Pacing and Ablation Technique Using Microelectrode for Pulmonary Vein Isolation Using a Local Impedance‐Guided Catheter
Open Access
- 27 January 2025
- journal article
- review article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 48 (2) , 216-223
Abstract
Background: The IntellaNav MiFi OI catheter (MiFi) is equipped with a sensor for local impedance (LI) monitoring and three mini‐electrodes. In this study, we investigated the target LI values for a successful pulmonary vein isolation (PVI) under the pacing and ablation technique using the MiFi catheter.Methods: Twenty‐seven patients underwent PVI using the MiFi catheter under mini electrode pacing from the MiFi catheter. The local impedance (LI) changes, generator impedance (GI) changes, and the time to capture loss were evaluated.Results: First‐pass isolations were obtained in 15 patients (57.7 %) for right PVs and in 22 patients (84.6 %) for left PVs. At gap sites, the impedance decrease was smaller than at non‐gap sites (non‐gap sites vs. gap sites; LI drop, 23.2 [±10.3] vs. 15.6 [±7.7] Ω, p < 0.0001; GI drop, 4.8 [±4.1] vs. 2.7 [3.9] Ω, p = 0.0026; %LI drop, −19.3 [±7.4] vs. −13.1 [±6.1] %, p < 0.0001; % GI drop, −5.1 [±4.2] vs. −2.9 [±4.2] %, p = 0.0020), suggesting that changes in impedance could be useful for predicting gaps. The cutoff values for predicting no gaps were identified as 15.0 Ω for the LI drop and −13.74% for the %LI drop.Conclusion: The LI showed greater changes than the GI and was also useful for predicting gaps. The cutoff values of 15.0 Ω for the LI drop and −13.74% for the %LI drop could predict conduction gaps. Under the monitoring of the LI, the pacing and ablation technique proved useful for PVI, even though the MiFi catheter does not have a CF sensor or ablation indices.Keywords
This publication has 22 references indexed in Scilit:
- Comparison of the electrophysiological properties of the pulmonary veins between paroxysmal and persistent atrial fibrillationJournal of Arrhythmia, 2023
- Does the same lesion index mean the same efficacy and safety profile: influence of the differential power, time, and contact force on the lesion size and steam pops under the same lesion indexJournal of Interventional Cardiac Electrophysiology, 2023
- Time dependency in the radiofrequency lesion formation for a local impedance guided catheter in an ex vivo experimental modelJournal of Arrhythmia, 2022
- Local catheter impedance drop during pulmonary vein isolation predicts acute conduction block in patients with paroxysmal atrial fibrillation: initial results of the LOCALIZE clinical trialEP Europace, 2021
- Relationship between the ablation index, lesion formation, and incidence of steam popsJournal of Arrhythmia, 2019
- In Vitro Validation of the Lesion Size Index to Predict Lesion Width and Depth After Irrigated Radiofrequency Ablation in a Porcine ModelJACC: Clinical Electrophysiology, 2017
- 2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation: Executive summaryJournal of Arrhythmia, 2017
- EFFICAS II: optimization of catheter contact force improves outcome of pulmonary vein isolation for paroxysmal atrial fibrillationEP Europace, 2015
- Circumferential pulmonary vein isolation and linear left atrial ablation as a single-catheter technique to achieve bidirectional conduction block: The pace-and-ablate approachHeart Rhythm, 2010
- Spontaneous Initiation of Atrial Fibrillation by Ectopic Beats Originating in the Pulmonary VeinsNew England Journal of Medicine, 1998