Abstract
The purpose of this study is to validate the accuracy of a Monte Carlo calculation model of a proton magnetic beam scanning delivery nozzle developed using the Geant4 toolkit. The Monte Carlo model was used to produce depth dose and lateral profiles, which were compared to data measured in the clinical scanning treatment nozzle at several energies. Comparisons were also made between measured and simulated off-axis profiles to test the accuracy of the model's magnetic steering. Comparison of the 80% distal dose fall-off values for the measured and simulated depth dose profiles agreed to within 1 mm for the beam energies evaluated. Agreement of the full width at half maximum values for the measured and simulated lateral fluence profiles was within 1.3 mm for all energies. The position of measured and simulated spot positions for the magnetically steered beams agreed to within 0.7 mm of each other. Based on these results, we found that the Geant4 Monte Carlo model of the beam scanning nozzle has the ability to accurately predict depth dose profiles, lateral profiles perpendicular to the beam axis and magnetic steering of a proton beam during beam scanning proton therapy.Keywords
This publication has 26 references indexed in Scilit:
- Shielding design for a laser-accelerated proton therapy systemPhysics in Medicine & Biology, 2007
- Cluster computing software for GATE simulationsMedical Physics, 2007
- Spread‐out Bragg peak and monitor units calculation with the Monte Carlo Code MCNPXMedical Physics, 2007
- Benchmarking analytical calculations of proton doses in heterogeneous matterMedical Physics, 2005
- Verification of the dose distributions with GEANT4 simulation for proton therapyIEEE Transactions on Nuclear Science, 2005
- Implementation of a new Monte Carlo-GEANT4 Simulation tool for the development of a proton therapy beam line and verification of the related dose distributionsIEEE Transactions on Nuclear Science, 2005
- Validation ofGEANT4, an object-oriented Monte Carlo toolkit, for simulations in medical physicsMedical Physics, 2004
- Geant4—a simulation toolkitNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- ROOT — An object oriented data analysis frameworkNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997
- Three-dimensional beam scanning for proton therapyNuclear Instruments and Methods in Physics Research, 1983