Biomechanical Measurements of Surgical Drilling Force and Torque in Human Versus Artificial Femurs
Authors:
- Michael OlsenMOMichael Olsen
- Meghan CrookshankMCMeghan Crookshank
- Ahmad GhasempoorAGAhmad Ghasempoor
- Rad ZderoRZRad Zdero
- 1 December 2012
- journal article
- Published by Wiley in Journal of Biomechanical Engineering
- Vol. 134 (12) , 124503
Abstract
Few experimental studies have examined surgical drilling in human bone, and no studies have inquired into this aspect for a popular commercially-available artificial bone used in biomechanical studies. Sixteen fresh-frozen human femurs and five artificial femurs were obtained. Cortical specimens were mounted into a clamping system equipped with a thrust force and torque transducer. Using a CNC machine, unicortical holes were drilled in each specimen at 1000 rpm, 1250 rpm, and 1500 rpm with a 3.2 mm diameter surgical drill bit. Feed rate was 120 mm/min. Statistical significance was set at p < 0.05. Force at increasing spindle speed (1000 rpm, 1250 rpm, and 1500 rpm), respectively, showed a range for human femurs (198.4 ± 14.2 N, 180.6 ± 14.0 N, and 176.3 ± 11.2 N) and artificial femurs (87.2 ± 19.3 N, 82.2 ± 11.2 N, and 75.7 ± 8.8 N). For human femurs, force at 1000 rpm was greater than at other speeds (p ≤ 0.018). For artificial femurs, there was no speed effect on force (p ≥ 0.991). Torque at increasing spindle speed (1000 rpm, 1250 rpm, and 1500 rpm), respectively, showed a range for human femurs (186.3 ± 16.9 N·mm, 157.8 ± 16.1 N·mm, and 140.2 ± 16.4 N·mm) and artificial femurs (67.2 ± 8.4 N·mm, 61.0 ± 2.9 N·mm, and 53.3 ± 2.9 N·mm). For human femurs, torque at 1000 rpm was greater than at other speeds (p < 0.001). For artificial femurs, there was no difference in torque for 1000 rpm versus higher speeds (p ≥ 0.228), and there was only a borderline difference between the higher speeds (p = 0.046). Concerning human versus artificial femurs, their behavior was different at every speed (force, p ≤ 0.001; torque, p < 0.001). For human specimens at 1500 rpm, force and torque were linearly correlated with standardized bone mineral density (sBMD) and the T-score used to clinically categorize bone quality (R ≥ 0.56), but there was poor correlation with age at all speeds (R ≤ 0.37). These artificial bones fail to replicate force and torque in human cortical bone during surgical drilling. To date, this is the largest series of human long bones biomechanically tested for surgical drilling.Keywords
This publication has 66 references indexed in Scilit:
- The Biomechanical Effect of Torsion on Humeral Shaft Repair Techniques for Completed Pathological FracturesJournal of Biomechanical Engineering, 2012
- Experimental investigations of forces and torque in conventional and ultrasonically-assisted drilling of cortical boneMedical Engineering & Physics, 2010
- The biomechanical analysis of three plating fixation systems for periprosthetic femoral fracture near the tip of a total hip arthroplastyJournal of Orthopaedic Surgery and Research, 2010
- A modified adaptive controller design for teleoperation systemsRobotics and Autonomous Systems, 2010
- Vertically Oriented Femoral Neck Fractures: Mechanical Analysis of Four Fixation TechniquesJournal of Orthopaedic Trauma, 2007
- Drilling of fiber reinforced plastics: A reviewJournal of Materials Processing Technology, 2007
- The Biomechanics of Human Femurs in Axial and Torsional Loading: Comparison of Finite Element Analysis, Human Cadaveric Femurs, and Synthetic FemursJournal of Biomechanical Engineering, 2006
- Microwave Drilling of BonesIEEE Transactions on Biomedical Engineering, 2006
- Influence of process parameters on cutting force and torque during drilling of glass–fiber polyester reinforced compositesComposite Structures, 2005
- Theoretical analysis on chatter vibration in drilling and its suppressionJournal of Materials Processing Technology, 2003