Dosimetric characterization of a synthetic single crystal diamond detector in clinical radiation therapy small photon beams
- 2 July 2012
- journal article
- radiation therapy-physics
- Published by Wiley in Medical Physics
- Vol. 39 (7Part1) , 4493-4501
Abstract
Purpose: To determine the potentialities of synthetic single crystal diamond Schottky diodes for accurate dose measurements in radiation therapy small photon beams. Methods: The dosimetric properties of a diamond‐based detector were assessed by comparison with a reference microionization chamber. The diamond device was operated at zero bias voltage under irradiation with high‐energy radiotherapic photon beams. The stability of the detector response and its dose and dose rate dependence were measured. Different square field sizes ranging from 1 × 1 cm2 to 10 × 10 cm2 were used during comparative dose distribution measurements by means of percentage depth dose curves (PDDs), lateral beam profiles, and output factors. The angular and temperature dependence of the diamond detector response were also studied. Results: The detector response shows a deviation from linearity of less than ±0.5% in the 0.01–7 Gy range and dose rate dependence below ±0.5% in the 1–6 Gy/min range. PDDs and output factors are in good agreement with those measured by the reference ionization chamber within 1%. No angular dependence is observed by rotating the detector along its axis, while ∼3.5% maximum difference is measured by varying the radiation incidence angle in the polar direction. The temperature dependence was investigated as well and a ±0.2% variation of the detector response is found in the 18–40 °C range. Conclusions: The obtained results indicate the investigated synthetic diamond‐based detector as a candidate for small field clinical radiation dosimetry in advanced radiation therapy techniques.Keywords
This publication has 39 references indexed in Scilit:
- Response to high-energy photons of PTW31014 PinPoint ion chamber with a central aluminum electrodeMedical Physics, 2008
- Total scatter factors of small beams: A multidetector and Monte Carlo studyMedical Physics, 2008
- Small fields: Nonequilibrium radiation dosimetryMedical Physics, 2007
- Measurement of output factors for small photon beamsMedical Physics, 2007
- The volume effect of detectors in the dosimetry of small fields used in IMRTMedical Physics, 2003
- The value of the PinPoint ion chamber for characterization of small field segments used in intensity-modulated radiotherapyPhysics in Medicine & Biology, 2000
- Comparative dosimetry in narrow high-energy photon beamsPhysics in Medicine & Biology, 2000
- Assessment of new small-field detectors against standard-field detectors for practical stereotactic beam data acquisitionPhysics in Medicine & Biology, 1999
- Comments on dose measurements for a narrow beam in radiosurgeryMedical Physics, 1993
- Measurements of dose distributions in small beams of 6 MV X-raysPhysics in Medicine & Biology, 1987