Abstract
A new type of an electrochemical flow cell designed for exchangeable ultramicroelectrodes (UMEs) is presented. The cell could be used with up to 8 different UMEs allowing one to use this cell for UME array measurements. The cell was characterized using both empirical data and finite element modeling. Au and Pt UMEs were used to detect catechol simultaneously, demonstrating the ability of this cell to be used as a sensory platform. The electroanalytical response of flow system was validated by direct measurements of H2O2 concentrations. Finally, stochastic nanoparticle measurements were successfully performed using the flow system, highlighting the future possibilities of this system for environmental monitoring.
Funding Information
  • Gouvernement du Canada | National Research Council Canada (Conseil national de recherches Canada) (RGPIN/05002-2014)

This publication has 42 references indexed in Scilit: