A numerical investigation on the effect of RF coil feed variability on global and local electromagnetic field exposure in human body models at 64 MHz
- 18 April 2017
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 79 (2) , 1135-1144
Abstract
Purpose: This study aims to investigate how the positions of the feeding sources of the transmit radiofrequency (RF) coil, field orientation direction with respect to the patient, and patient dimensions affect the global and local electromagnetic exposure in human body models.Methods: Three RF coil models were implemented, namely a specific two‐source (S2) feed and two multisource feed configurations: generic 32‐source (G32) and hybrid 16‐source (H16). Thirty‐two feeding conditions were studied for the S2, whereas two were studied for the G32 and H16. The study was performed using five human body models. Additionally, for two of the body models, the case of a partially implanted lead was evaluated.Results: The results showed an overall variation due to coil feeding conditions of the whole‐body specific absorption rate (SAR) of less than 20%, but deviations up to 98% of the magnitude of the electric field tangential to a possible lead path. For the analysis with the partially implanted lead, a variation of local SAR at the tip of the lead of up to 60% was observed with respect to feed position and field orientation direction.Conclusion: The results of this study suggest that specific information about feed position and field orientation direction must be considered for an accurate evaluation of patient exposure. Magn Reson Med 79:1135–1144, 2018. © 2017 International Society for Magnetic Resonance in Medicine.Keywords
This publication has 18 references indexed in Scilit:
- Evaluation of the RF heating of a generic deep brain stimulator exposed in 1.5 T magnetic resonance scannersBioelectromagnetics, 2012
- Pacemaker lead tip heating in abandoned and pacemaker‐attached leads at 1.5 tesla MRIJournal of Magnetic Resonance Imaging, 2011
- The Virtual Family—development of surface-based anatomical models of two adults and two children for dosimetric simulationsPhysics in Medicine & Biology, 2009
- Complexity of MRI induced heating on metallic leads: Experimental measurements of 374 configurationsBioMedical Engineering OnLine, 2008
- Calculation of MRI‐induced heating of an implanted medical lead wire with an electric field transfer functionJournal of Magnetic Resonance Imaging, 2007
- Calculations ofB 1 distribution, specific energy absorption rate, and intrinsic signal-to-noise ratio for a body-size birdcage coil loaded with different human subjects at 64 and 128 MHzApplied Magnetic Resonance, 2005
- RF heating due to conductive wires during MRI depends on the phase distribution of the transmit fieldMagnetic Resonance in Medicine, 2002
- Invited. Interventional and intraoperative MRI: A general overview of the fieldJournal of Magnetic Resonance Imaging, 1998
- The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHzPhysics in Medicine & Biology, 1996
- Analysis of the numerical error caused by the stair-stepped approximation of a conducting boundary in FDTD simulations of electromagnetic phenomenaIEEE Transactions on Antennas and Propagation, 1991