Axial Loads on Implant-Supported Partial Fixed Prostheses for External and Internal Hex Connections and Machined and Plastic Copings: Strain Gauge Analysis
- 1 April 2015
- journal article
- Published by American Academy of Implant Dentistry in Journal of Oral Implantology
- Vol. 41 (2) , 149-154
Abstract
The aim of this in vitro study was to use strain gauge (SG) analysis to compare the effects of the implant-abutment joint, the coping, and the location of load on strain distribution in the bone around implants supporting 3-unit fixed partial prostheses. Three external hexagon (EH) implants and 3 internal hexagon (IH) implants were inserted into 2 polyurethane blocks. Microunit abutments were screwed onto their respective implant groups. Machined cobalt-chromium copings and plastic copings were screwed onto the abutments, which received standard wax patterns. The wax patterns were cast in a cobalt-chromium alloy (n = 5): group 1 = EH/machined. group 2 = EH/plastic, group 3 = IH/machined, and group 4 = IH/plastic. Four SGs were bonded onto the surface of the block tangentially to the implants. Each metallic structure was screwed onto the abutments and an axial load of 30 kg was applied at 5 predetermined points. The magnitude of microstrain on each SG was recorded in units of microstrain (μɛ). The data were analyzed using 3-factor repeated measures analysis of variance and a Tukey test (α = 0.05). The results showed statistically significant differences for the type of implant-abutment joint, loading point, and interaction at the implant-abutment joint/loading point. The IH connection showed higher microstrain values than the EH connection. It was concluded that the type of coping did not interfere in the magnitude of microstrain, but the implant/abutment joint and axial loading location influenced this magnitude.Keywords
This publication has 20 references indexed in Scilit:
- Machined and plastic copings in three-element prostheses with different types of implantabutment joints: a strain gauge comparative analysisJournal of Applied Oral Science, 2010
- External Hexagon and Internal Hexagon in Straight and Offset Implant Placement: Strain Gauge AnalysisImplant Dentistry, 2009
- The Staggered Installation of Dental Implants and Its Effect on Bone StressesClinical Implant Dentistry and Related Research, 2007
- Influence of forces on peri‐implant boneClinical Oral Implants Research, 2006
- Loading of bone surrounding implants through three‐unit fixed partial denture fixation: a finite‐element analysis based on in vitro and in vivo strain measurementsClinical Oral Implants Research, 2006
- Strain Situation after Fixation of Three-Unit Ceramic Veneered Implant SuperstructuresImplant Dentistry, 2005
- Cement fixation and screw retention: parameters of passive fitClinical Oral Implants Research, 2004
- The influence of functional forces on the biomechanics of implant-supported prostheses—a reviewJournal of Dentistry, 2002
- Brånemark System® and ITI Dental Implant System® for treatment of mandibular edentulismClinical Oral Implants Research, 2001
- Histological evaluation of peri-implant bone at implants subjected to occlusal overload or plaque accumulationClinical Oral Implants Research, 1997