Development and Initial Evaluation of a Cost-Effective Force Sensor for Ureteroscopic Application
Authors:
- Bruce M. GaoBGBruce M. Gao
- Jacob C. TsaiJTJacob C. Tsai
- Andrei D. CumpanasACAndrei D. Cumpanas
- Jaime Altamirano-VillarroelJAJaime Altamirano-Villarroel
- Seyedamirvala SaadatSSSeyedamirvala Saadat
- Victor PhamVPVictor Pham
- Evan GrohsEGEvan Grohs
- Yi Xi WuYWYi Xi Wu
- Zachary E. TanoZTZachary E. Tano
- Sohrab N. AliSASohrab N. Ali
- Pengbo JiangPJPengbo Jiang
- Roshan M. PatelRPRoshan M. Patel
- 1 October 2024
- journal article
- research article
- Published by SAGE Publications in Journal of Endourology
- Vol. 38 (10) , 1075-1081
Abstract
Purpose: Retrograde intrarenal surgery is the gold-standard treatment for most kidney stones. During ureteroscopy, ureteral access sheath insertion at forces greater than 8.0 Newtons (N) risks high-grade ureteral injury. To monitor force, our institution utilizes a unique, Bluetooth-equipped device (i.e., the University of California—Irvine Force Sensor). Given the unique nature of the force sensor, we sought to develop an inexpensive and accessible force sensor based on Boyle’s law and the specific amount of force required to compress an occluded 1.0 mL syringe. Materials and Methods: We evaluated three brands of 1.0 mL syringes. After setting the plunger at 1.0 mL, the syringe was occluded, and the syringe plunger was compressed. The syringe volume was recorded when the applied force on the plunger reached 4.0 N, 6.0 N, and 8.0 N. Multiple trials were performed to assess reliability and reproducibility. A method for applying this clinically was also developed. Results: The precise force thresholds identified for a 1.0 mL Luer-Lok™ Syringe (Becton Dickinson, Franklin Lakes, NJ) were 0.30 mL for 4.00 N, 0.20 mL for 6.00 N, and 0.15 mL for 8.00 N. The 1.0 mL Tuberculin Syringe and 1.0 mL Luer Slip Syringe were less precise, but compression from 1.0 to 0.40 mL, 0.25 mL, and 0.20 mL corresponded to force sensor readings that did not exceed 4.00 N, 6.00 N, and 8.00 N, respectively. Conclusions: Based on volume changes, 4.00 N, 6.00 N, and 8.00 N of force can be reliably and reproducibly achieved using an occluded 1.0 mL syringe.Keywords
This publication has 9 references indexed in Scilit:
- MP35-03 SAFE PASSAGE OF 16 FR URETERAL ACCESS SHEATHSJournal of Urology, 2023
- Determining the Safety Threshold for the Passage of a Ureteral Access Sheath in Clinical Practice Using a Purpose-Built Force SensorJournal of Urology, 2021
- Ureteral Wall Injury with Ureteral Access Sheaths: A Randomized Prospective TrialJournal of Endourology, 2020
- Ureteral Access Sheath Deployment: How Much Force Is Too Much? Initial Studies with a Novel Ureteral Access Sheath Force Sensor in the Porcine UreterJournal of Endourology, 2019
- Teaching Boyle’s Law and Charles’ Law through Experiments that Use Novel, Inexpensive Equipment Yielding Accurate ResultsJournal of Chemical Education, 2018
- Surgical Management of Stones: American Urological Association/Endourological Society Guideline, PART IIJournal of Urology, 2016
- Differences in renal stone treatment and outcomes for patients treated either with or without the support of a ureteral access sheath: The Clinical Research Office of the Endourological Society Ureteroscopy Global StudyWorld Journal of Urology, 2015
- Systemic Inflammatory Response Syndrome After Flexible Ureteroscopic Lithotripsy: A Study of Risk FactorsJournal of Endourology, 2015
- Characterization of intrapelvic pressure during ureteropyeloscopy with ureteral access sheathsUrology, 2003