Skin and build up dose determination for a 2.5 MV medical linear accelerator imaging beam
- 9 September 2019
- journal article
- research article
- Published by Springer Nature in Australasian Physics & Engineering Sciences in Medicine
- Vol. 42 (4) , 1177-1181
Abstract
The 2.5 MV Imaging beam produced by a Varian TrueBeam linear accelerator produces a dose build up effect at the beam entrance similar to other high energy photon beams. The surface dose values were found to range from 39% of maximum dose at a 5 cm × 5 cm field size up to 69% of maximum at a 40 cm × 40 cm field. The depth of maximum dose deposition was found to range from 5 mm at smaller field sizes to 4 mm at larger field sizes. Whilst large absorbed doses will not be delivered utilizing these beams, the data provided will allow the medical physics community to assess and estimate doses to patient’s skin and subcutaneous tissue from low energy MV imaging beams.Keywords
This publication has 21 references indexed in Scilit:
- Validating kQ =1.0 assumption in TG51 with PTW 30013 farmer chamber for Varian TrueBeam's 2.5 MV imaging beamJournal of Applied Clinical Medical Physics, 2018
- Characterization of a 2.5 MV inline portal imaging beamJournal of Applied Clinical Medical Physics, 2016
- Characterization and evaluation of 2.5 MV electronic portal imaging for accurate localization of intra‐ and extracranial stereotactic radiosurgeryJournal of Applied Clinical Medical Physics, 2016
- SU‐E‐J‐147: Quantitative Analysis of 2.5 MV Portal Imaging Performance Compared to KV and 6MV Portal Imaging On the Novel Edge LINACMedical Physics, 2014
- Surface Dose Investigation of the Flattening Filter-Free Photon BeamsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012
- A study of surface dosimetry for breast cancer radiotherapy treatments using Gafchromic EBT2 filmJournal of Applied Clinical Medical Physics, 2012
- Assessment of skin dose for breast chest wall radiotherapy as a function of bolus materialPhysics in Medicine & Biology, 2008
- Accurate skin dose measurements using radiochromic film in clinical applicationsMedical Physics, 2006
- X‐ray surface dose measurements using TLD extrapolationMedical Physics, 1993
- Revised Code of Practice for the dosimetry of 2 to 35 MV x-ray, and of caesium-137 and cobalt-60 gamma-ray beamsPhysics in Medicine & Biology, 1983