Dosimetric validation of the Acuros XB Advanced Dose Calculation algorithm: fundamental characterization in water
Authors:
- Antonella FogliataAFAntonella Fogliata
- Giorgia NicoliniGNGiorgia Nicolini
- Alessandro ClivioACAlessandro Clivio
- Eugenio VanettiEVEugenio Vanetti
- Pietro MancosuPMPietro Mancosu
- Luca CozziLCLuca Cozzi
- 1 March 2011
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 56 (6) , 1879-1904
Abstract
A new algorithm, Acuros® XB Advanced Dose Calculation, has been introduced by Varian Medical Systems in the Eclipse planning system for photon dose calculation in external radiotherapy. Acuros XB is based on the solution of the linear Boltzmann transport equation (LBTE). The LBTE describes the macroscopic behaviour of radiation particles as they travel through and interact with matter. The implementation of Acuros XB in Eclipse has not been assessed; therefore, it is necessary to perform these pre-clinical validation tests to determine its accuracy. This paper summarizes the results of comparisons of Acuros XB calculations against measurements and calculations performed with a previously validated dose calculation algorithm, the Anisotropic Analytical Algorithm (AAA). The tasks addressed in this paper are limited to the fundamental characterization of Acuros XB in water for simple geometries. Validation was carried out for four different beams: 6 and 15 MV beams from a Varian Clinac 2100 iX, and 6 and 10 MV 'flattening filter free' (FFF) beams from a TrueBeam linear accelerator. The TrueBeam FFF are new beams recently introduced in clinical practice on general purpose linear accelerators and have not been previously reported on. Results indicate that Acuros XB accurately reproduces measured and calculated (with AAA) data and only small deviations were observed for all the investigated quantities. In general, the overall degree of accuracy for Acuros XB in simple geometries can be stated to be within 1% for open beams and within 2% for mechanical wedges. The basic validation of the Acuros XB algorithm was therefore considered satisfactory for both conventional photon beams as well as for FFF beams of new generation linacs such as the Varian TrueBeam.Keywords
This publication has 24 references indexed in Scilit:
- Dosimetric verification of the anisotropic analytical algorithm in lung equivalent heterogeneities with and without bone equivalent heterogeneitiesMedical Physics, 2010
- The effect of different lung densities on the accuracy of various radiotherapy dose calculation methods: Implications for tumour coverageRadiotherapy and Oncology, 2009
- Optimization of deterministic transport parameters for the calculation of the dose distribution around a high dose‐rate brachytherapy sourceMedical Physics, 2008
- Experimental validation of the Eclipse AAA algorithmJournal of Applied Clinical Medical Physics, 2007
- On the dosimetric behaviour of photon dose calculation algorithms in the presence of simple geometric heterogeneities: comparison with Monte Carlo calculationsPhysics in Medicine & Biology, 2007
- Dosimetric verification of the anisotropic analytical algorithm for radiotherapy treatment planningRadiotherapy and Oncology, 2006
- Comparison of a finite-element multigroup discrete-ordinates code with Monte Carlo for radiotherapy calculationsPhysics in Medicine & Biology, 2006
- Dosimetric validation of the anisotropic analytical algorithm for photon dose calculation: fundamental characterization in waterPhysics in Medicine & Biology, 2006
- American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planningMedical Physics, 1998
- Dosimetric Precision Requirements in Radiation TherapyActa Radiologica: Oncology, 1984