Evaluation of the Aesculon cardiac output monitor by subxiphoidal Doppler flow measurement in children with congenital heart defects
Authors:
- Maren TomaskeMTMaren Tomaske
- Walter KnirschWKWalter Knirsch
- Oliver KretschmarOKOliver Kretschmar
- Christian BalmerCBChristian Balmer
- Katja WoitzekKWKatja Woitzek
- Achim SchmitzASAchim Schmitz
- Urs BauersfeldUBUrs Bauersfeld
- Markus WeissMWMarkus Weiss
- 1 May 2009
- journal article
- research article
- Published by Wolters Kluwer Health in European Journal of Anaesthesiology
- Vol. 26 (5) , 412-415
Abstract
To evaluate the noninvasive electrical velocimetry (Aesculon) monitor for cardiac output (CO) by subxiphoidal Doppler flow measurement in children. CO was determined at the end of diagnostic or interventional cardiac catheterization for congenital heart defects. Standard ECG surface electrodes were attached in a vertical direction to the patients' left middle and lower neck, and lower thorax at the level of the heart and xiphoid process. Aesculon CO data were compared with a simultaneously measured CO by the subxiphoidal Doppler flow measurement technique. For each patient, measurements were repeated three times within 5 min. Whitney U-test, simple regression and Bland-Altman analysis were performed to compare CO values obtained by the two techniques. Data are given as range (median). A total of 36 children aged 5.7 (0.5-16.0) years were investigated. CO values obtained by Aesculon monitor [0.55-5.58 (2.62) l min] and subxiphoidal Doppler flow measurements [0.62-6.27 (3.05) l min] differed significantly between both methods (P = 0.04). Simple regression analysis revealed moderate correlation between CO values obtained from the two techniques (r = 0.5544, P < 0.001). Bias between the two methods was 0.31 l min with a precision of 1.92 l min. We conclude that electrical velocimetry using the Aesculon monitor does not reliably reflect absolute CO values as compared with subxiphoidal Doppler flow measurement.Keywords
This publication has 16 references indexed in Scilit:
- Cardiac output measurement in children: comparison of Aesculon® cardiac output monitor and thermodilutionBritish Journal of Anaesthesia, 2008
- Bernstein-Osypka stroke volume equation for impedance cardiography: citation correction.Intensive Care Medicine, 2007
- Iatrogenic pulmonary artery ruptureCurrent Opinion in Anaesthesiology, 2007
- Cardiac index monitoring by pulse contour analysis and thermodilution after pediatric cardiac surgeryThe Journal of Thoracic and Cardiovascular Surgery, 2007
- RETRACTED ARTICLE: Noninvasive assessment of cardiac output using thoracic electrical bioimpedance in hemodynamically stable and unstable patients after cardiac surgery: a comparison with pulmonary artery thermodilutionIntensive Care Medicine, 2006
- Comparison of electrical velocimetry and transoesophageal Doppler echocardiography for measuring stroke volume and cardiac output † †Declaration of interest. This study was supported by Osypka Medical GmbH, Berlin, Germany.British Journal of Anaesthesia, 2005
- Stroke volume equation for impedance cardiographyMedical & Biological Engineering & Computing, 2005
- A new stroke volume equation for thoracic electrical bioimpedanceCritical Care Medicine, 1986
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Complications of vascular catheterization in critically ill childrenCritical Care Medicine, 1984