Exogenous interferences with Jaffe creatinine assays: addition of sodium dodecyl sulfate to reagent eliminates bilirubin and total protein interference with Jaffe methods
Authors:
- 1 January 2010
- journal article
- research article
- Published by Wiley in Journal of Clinical Laboratory Analysis
- Vol. 24 (3) , 123-133
Abstract
Background: The study evaluated the impact of interferences on the analytical specificity of three commercial and commonly used creatinine methods (two Jaffe and one enzymatic). Methods: Manufacturer creatinine methods plus modified methods were tested with the following interferences: spiking serum with bilirubin, albumin, glucose, hemoglobin and lipid, and patient sera with maximum concentrations of bilirubin, 1,090 µmol/l and protein, 117.8 g/l. Results: Hemoglobin, 7.5 g/l and lipaemic with triglyceride concentration of 6.27 mmol/l, did not interfere with all assays. Glucose >33.3 mmol/l increased creatinine recovery for Dimension method. Samples spiked with bilirubin imparted a negative bias for Dimension and Architect methods but imparted a positive bias for Vitros assay. However, using patient sera, negative bias with bilirubin was found for all methods, from which Architect method gave the highest effect (R2=0.861), followed by Vitros (R2=0.239) and Dimension (R2=0.163). Protein provided the positive bias for all creatinine measurements that increased with increasing concentration (R2 ranging from 0.104 to 0.182, PConclusions: Endogenous interferences continue to plague creatinine accuracy measurement in both Jaffe and enzymatic methods, and consequentially the estimated glomerular filtration rate. The addition of SDS to the alkaline‐pirate reagent was shown to be effective in reducing bilirubin and protein interferences. J. Clin. Lab. Anal. 24:123–133, 2010.Keywords
This publication has 30 references indexed in Scilit:
- Use of Single-Value Protein Compensation of the Jaffe Creatinine Assay Contributes to Clinically Significant Inaccuracy in ResultsClinical Chemistry, 2008
- Different Methods of Creatinine Measurement Significantly Affect MELD ScoresLiver Transplantation, 2007
- Does Bilirubin Cause Interference in Roche Creatinine Methods?Clinical Chemistry, 2007
- Does the ID-MS traceable MDRD equation work and is it suitable for use with compensated Jaffé and enzymatic creatinine assays?Nephrology Dialysis Transplantation, 2006
- Recommendations for Improving Serum Creatinine Measurement: A Report from the Laboratory Working Group of the National Kidney Disease Education ProgramClinical Chemistry, 2006
- Chronic Kidney Disease in the Elderly — How to Assess RiskNew England Journal of Medicine, 2005
- Determination of reference intervals for serum creatinine, creatinine excretion and creatinine clearance with an enzymatic and a modified Jaffé methodClinica Chimica Acta, 2004
- Comparison of techniques for minimizing interference of bilirubin on serum creatinine determined by the kinetic Jaffé reactionJournal of Clinical Laboratory Analysis, 1994
- Prediction of glomerular filtration rate from serum creatinine concentration in advanced chronic renal failureKidney International, 1993
- Prediction of Creatinine Clearance from Serum CreatinineNephron, 1976