A Parametric Analysis of Endovascular Stent Geometry Manipulation on Radial Force Performance
- 1 October 2020
- journal article
- research article
- Published by ASME International in Journal of Medical Devices
- Vol. 14 (4)
Abstract
Stents were manufactured to investigate the impact of altering stent design characteristics (leg length, bend angle, bend radius, and wire diameter) on radial force generation. Results from this design parameter study showed that leg length, bend angle, and wire diameter have a statistically significant impact on radial force generation, while lesser changes in bend radius did not (1.00 mm versus the original 0.794 mm). However, a larger variation of this parameter (1.588 mm) was statistically significant. Results gathered for all parameters were used in the creation of a prototypal software. Using input values of patient-specific arterial diameter and compliance, as well as stent design characteristic dimensions, this program has been developed to predict stent radial force at varying levels of expansion.Keywords
Funding Information
- the Natural Sciences and Engineering Research Council of Canada (Nos. CDEPJ 245445-15 and RGPIN-2015-05639)
This publication has 11 references indexed in Scilit:
- Endovascular Repair of Complicated Acute Type-B Aortic Dissection with Stentgraft: Early and Mid-term ResultsEuropean Journal of Vascular and Endovascular Surgery, 2011
- Successful Treatment of Superior Vena Cava Rupture with Placement of a Covered Stent: A Report of Two CasesCardioVascular and Interventional Radiology, 2011
- The Mechanical Properties of Endovascular Stents: An In Vitro AssessmentCardiovascular Engineering, 2010
- AAA Stent–Grafts: Past Problems and Future ProspectsAnnals of Biomedical Engineering, 2010
- A Review of the In Vivo and In Vitro Biomechanical Behavior and Performance of Postoperative Abdominal Aortic Aneurysms and Implanted Stent-GraftsJournal of Endovascular Therapy, 2008
- Endovascular Management of Thoracic DissectionsCirculation, 2008
- Development of endovascular stent-graftsProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2007
- Stent-graft design: the good, the bad and the uglyCardiovascular Surgery, 2001
- Characterization of Radial Forces in Z StentsInvestigative Radiology, 2001
- Incidence and risk factors of late rupture, conversion, and death after endovascular repair of infrarenal aortic aneurysms: The EUROSTAR experienceJournal of Vascular Surgery, 2000