The Microsoft Research Aurora Project: Important Findings on Cuffless Blood Pressure Measurement
Top Cited Papers
- 1 March 2023
- journal article
- review article
- Published by Wolters Kluwer Health in Hypertension
- Vol. 80 (3) , 534-540
Abstract
Conventional blood pressure (BP) measurement devices based on an inflatable cuff only provide a narrow view of the continuous BP profile. Cuffless BP measuring technologies could permit numerous BP readings throughout daily life and thereby considerably improve the assessment and management of hypertension. Several wearable cuffless BP devices based on pulse wave analysis (applied to a photoplethysmography or tonometry waveform) with or without use of pulse arrival time are now available on the market. The key question is: Can these devices provide accurate measurement of BP? Microsoft Research recently published a complex article describing perhaps the most important and highest resource project to date (Aurora Project) on assessing the accuracy of several pulse wave analysis and pulse wave analysis-pulse arrival time devices. The overall results from 1125 participants were clear-cut negative. The present article motivates and describes emerging cuffless BP devices and then summarizes the Aurora Project. The study methodology and findings are next discussed in the context of regulatory-cleared devices, physiology, and related studies, and the study strengths and limitations are pinpointed thereafter. Finally, the implications of the Aurora Project are briefly stated and recommendations for future work are offered to finally realize the considerable potential of cuffless BP measurement in health care.Keywords
This publication has 22 references indexed in Scilit:
- 2021 European Society of Hypertension practice guidelines for office and out-of-office blood pressure measurementJournal Of Hypertension, 2021
- Blood pressure from the optical Aktiia Bracelet: a 1-month validation study using an extended ISO81060-2 protocol adapted for a cuffless wrist deviceBlood Pressure Monitoring, 2021
- Nocturnal blood pressure measured by home devicesJournal Of Hypertension, 2019
- A Universal Standard for the Validation of Blood Pressure Measuring DevicesHypertension, 2018
- Comparison of noninvasive pulse transit time estimates as markers of blood pressure using invasive pulse transit time measurements as a referencePhysiological Reports, 2016
- Toward Ubiquitous Blood Pressure Monitoring via Pulse Transit Time: Theory and PracticeIEEE Transactions on Biomedical Engineering, 2015
- The use of ambulatory tonometric radial arterial wave capture to measure ambulatory blood pressure: the validation of a novel wrist-bound device in adultsJournal of Human Hypertension, 2007
- Reactive rise in blood pressure upon cuff inflation: cuff inflation at the arm causes a greater rise in pressure than at the wrist in hypertensive patientsBlood Pressure Monitoring, 2007
- Wearable, Cuff-less PPG-Based Blood Pressure Monitor with Novel Height Sensor2006
- Effects of cuff inflation on self-recorded blood pressureThe Lancet, 1990