Glucose measurements with accu check inform II versus hexokinase plasma method during surgery under general anesthesia, an observational cohort study
- 9 September 2023
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 38 (2) , 479-485
Abstract
Purpose Limited research exists on translation of in-vitro glucose measurement interfering compounds to the in-vivo situation. We investigated whether Point-of-Care glucose measurements by Accu Chek Inform II (ACI II) were accurate to monitor glucose concentrations during surgery with general anesthesia by comparing with the reference laboratory hexokinase plasma glucose test. Method Patients undergoing surgery with general anesthesia were included. Anesthesia was maintained with either Sevoflurane or Total intravenous anesthesia (TIVA). Prior to and after induction, blood glucose was measured with ACI II and the hexokinase test. Bland-Altman analysis was performed to assess method agreement. Subgroup analyses on glucose measurement differences per type of maintenance anesthesia were performed. Results Thirty-nine patients were included, and 78 measurements were performed. All paired measurements had clinically acceptable agreement with a percentage error of 10.0% (95% CI 8.0 to 11.9). The mean difference (95% limits of agreement) between ACI II and hexokinase for all measurements was 0.0 mmol/L (-0.7 to 0.7 mmol/L). Before induction (n = 39), mean difference was -0.1 mmol/L (-0.6 to 0.4 mmol/L), and after induction (n = 39), mean difference was 0.1 mmol/L (-0.8 to 0.9 mmol/L). Further investigation showed the difference varied per test for patients receiving Sevoflurane compared to patients receiving TIVA (-0.2 ± 0.4 mmol/L vs. 0.4 ± 0.3 mmol/L, p < 0.001). Before and after induction, the difference between ACI II and hexokinase measurements increased for patients receiving Sevoflurane compared to patients receiving TIVA (0.4 ± 0.4 mmol/L vs. -0.4 ± 0.3 mmol/L, p < 0.001). Conclusion The agreement between glucose measurements using ACI II and the reference laboratory hexokinase test was clinically acceptable with a percentage error of 10.0% (95% CI 8.0 to 11.9). The use of TIVA may negatively affect the measurement performance of the ACI II.Keywords
This publication has 14 references indexed in Scilit:
- Bedside Glucose Monitoring—Is it Safe? A New, Regulatory-Compliant Risk Assessment Evaluation Protocol in Critically Ill Patient Care Settings*Critical Care Medicine, 2017
- Technical and clinical validation of the Greiner FC-Mix glycaemia tubeClinical Chemistry and Laboratory Medicine (CCLM), 2017
- Effects of various types of anesthesia on hemodynamics, cardiac function, and glucose and lipid metabolism in ratsAmerican Journal of Physiology-Heart and Circulatory Physiology, 2016
- Acute effects of general anesthesia with propofol, pentobarbital or isoflurane plus propofol on plasma metabolites and hormones in adult pigs1Journal of Animal Science, 2016
- Stress-induced hyperglycemia is associated with higher mortality in severe traumatic brain injuryJournal of Trauma and Acute Care Surgery, 2015
- The Accuracy of Point-of-Care Glucose MeasurementsJournal of Diabetes Science and Technology, 2012
- The Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) statement: guidelines for reporting observational studiesThe Lancet, 2007
- Agreement Between Methods of Measurement with Multiple Observations Per IndividualJournal of Biopharmaceutical Statistics, 2007
- The stress response to trauma and surgeryBritish Journal of Anaesthesia, 2000
- Effects of Drugs on Glucose Measurements With Handheld Glucose Meters and a Portable Glucose AnalyzerAmerican Journal of Clinical Pathology, 2000