The rebreathing characteristics of the Hudson oxygen mask
Authors:
- 1 March 1993
- journal article
- Published by Wiley in Anaesthesia
- Vol. 48 (3) , 239-242
Abstract
Summary: A lung model was used to study the effect of varying minute ventilation, oxygen flow and inspiratory: expiratory ratio on the rebreathing chracteristics of the Hudson oxygen mask. The onset of rebreathing was detected using two different sets of criteria; those defined by Kain et al. and Humphrey. The onset of rebreathing was also quantified in terms of the inspired volume of carbon dioxide and increases in deadspace. The magnitude of rebreathing was increased by reducing oxygen inflow and increasing minute ventilation. Variation in I:E ratio between 1:1 and 1:4 had little effect on rebreathing. Using the clinical criteria of Kain et al, the onset of rebreathing was detected when oxygen inflow/minute ventilation ratio was reduced below 0.78 on average. With the other criterion, rebreathing occurred when the ratio was 0.46.Keywords
This publication has 9 references indexed in Scilit:
- Rebreathing during oxygen treatment with face maskActa Anaesthesiologica Scandinavica, 1991
- A comparison of anaesthetic breathing systems during spontaneous ventilationAnaesthesia, 1989
- The Lack, Magill and Bain Anaesthetic Breathing Systems: A Direct Comparison in Spontaneously-Breathing Anaesthetized AdultsJournal of the Royal Society of Medicine, 1982
- Breath-by-breath VCO2 and VO2 required compensation for transport delay and dynamic responseJournal of Applied Physiology, 1982
- Respiratory Waveform and Rebreathing in T-Piece CircuitsAnesthesiology, 1980
- Fresh Gas Original Investigations of the Magill CircuitAnesthesiology, 1968
- THE PERFORMANCE OF NASAL CATHETERS AND A FACE MASK IN OXYGEN THERAPYThe Lancet, 1967
- A METHOD OF CONTROLLED OXYGEN ADMINISTRATION WHICH REDUCES THE RISK OF CARBON-DIOXIDE RETENTIONThe Lancet, 1960
- Reassessment of Value of Oxygen Masks that Permit RebreathingBMJ, 1956