DRY ELECTRODE MATERIAL TESTING AND GAIN ANALYSIS FOR SINGLE ARM ECG SYSTEM
Authors:
- 1 June 2017
- journal article
- Published by National Taiwan University in Biomedical Engineering: Applications, Basis and Communications
- Vol. 29 (03)
Abstract
Electrocardiography (ECG) is a medical diagnostic procedure used to record the electrical activity of the heart and display it as a waveform. For picking the ECG waveform, wet and dry electrodes can be utilized. Lots of dry electrode related studies have been carried out in different parts of the world. This paper focuses on the acquisition of ECG from the left arm and right arm separately and comparing the waveforms and gain required for the system. In addition to that, different dry electrode materials and dimensions are used to check the acquisition of ECG waveform from the single arm. Dry electrodes employed in this study include copper, brass, phosphor bronze and nickel silver. The various dimensions used in this study are 3cm×3cm×0.5mm, 2.5cm×2.7cm×0.5mm, 2.5cm×2.5cm×0.5mm, 2cm×3cm×0.5mm, 2cm×2cm×0.5mm and 1cm×3cm×0.5mm. A total of five healthy subjects are used in this study for the acquisition of ECG. The results are very much promising in that the gain required for the left arm ECG system is less...Keywords
This publication has 5 references indexed in Scilit:
- DESIGNING OF A SINGLE ARM SINGLE LEAD ECG SYSTEM FOR WET AND DRY ELECTRODE: A COMPARISON WITH TRADITIONAL SYSTEMBiomedical Engineering: Applications, Basis and Communications, 2016
- ECG monitoring leads and special leadsIndian Pacing and Electrophysiology Journal, 2016
- Soft, Comfortable Polymer Dry Electrodes for High Quality ECG and EEG RecordingSensors, 2014
- Dry electrodes for electrocardiographyPhysiological Measurement, 2013
- Dry-Contact and Noncontact Biopotential Electrodes: Methodological ReviewIEEE Reviews in Biomedical Engineering, 2010