Cortex Distraction Forces Caused by the Insertion of External Fixator Pins
Authors:
- Peter Brian MacFarlane ThomasPTPeter Brian MacFarlane Thomas
- Stephan VerborgSVStephan Verborg
- Peter Jan OgrodnikPOPeter Jan Ogrodnik
- 1 September 2001
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Orthopaedic Trauma
- Vol. 15 (7) , 507-512
Abstract
Insertion of a bicortical threaded external fixator pin can cause bone damage. An unexpected crack can propagate along the bone when the pin touches the far cortex. The objective of this study was to investigate whether drilling and inserting bicortical pins into the shafts of long bones can cause large distraction forces to be generated between the cortices. Two flat samples of bone or bone substitute (Tufnol) were mounted parallel at each end of a specially designed force measurement column. Three common pin designs were inserted into the samples: tapered, self-threading and self-drilling, and self-threading pins. The axial thrust and the cortex distraction force between the two samples were measured as the pins were inserted. High distraction forces were measured between the cortices for all pin designs, typically 700 to 1,000 Newtons. The clinical significance of these findings is that the insertion of bicortical pins of the designs tested may generate a large distraction force between the cortices. This may drive a fracture apart or propagate an undisplaced fracture. Therefore, care must be taken to ensure that bicortical pins are inserted in locations free of local defects.Keywords
This publication has 12 references indexed in Scilit:
- The Ilizarov fixator in trauma: a 10-year experienceJournal of Orthopaedic Science, 2000
- Pin-tract complications in external fixation of fractures of the distal radiusActa Orthopaedica, 1999
- 134 tibial shaft fractures managed with the Dynamic Axial FixatorActa Orthopaedica, 1995
- Analysis of the External Fixator Pin-Bone InterfaceClinical Orthopaedics and Related Research, 1993
- Devenir à moyen terme de 109 ostéonécroses aseptiques de la tête fémorale : Le forage est-il encore d'actualité?European Journal of Orthopaedic Surgery & Traumatology, 1992
- Design and testing of external fixator bone screwsJournal of Biomedical Engineering, 1990
- Bending Stiffness of Unilateral and Bilateral Fixator FramesClinical Orthopaedics and Related Research, 1983
- Research on the mechanical impedance of human bone by a drilling testJournal of Biomechanics, 1981
- The Effect of Movement on the Holding Power of Screws in BoneClinical Orthopaedics and Related Research, 1975
- Temperatures Measured in Human Cortical Bone when DrillingJournal of Bone and Joint Surgery, 1972