Micro-CT Evaluation of Microgaps at Implant-Abutment Connection
Open Access
Abstract
The assessment of microgaps at the implant–abutment interface is an important factor that may influence clinical success. Thus, the aim of this study was to evaluate the size of microgaps between prefabricated and customised abutments (Astra Tech, Dentsply, York, PA, USA; Apollo Implants Components, Pabianice, Poland) mounted on a standard implant. The measurement of the microgap was performed using micro-computed tomography (MCT). Due to 15-degree rotation of samples, 24 microsections were obtained. Scans were performed at four levels established at the interface between the abutment and the implant neck. Moreover, the volume of the microgap was evaluated. The size of the microgap at all measured levels varied from 0.1 to 3.7 µm for Astra and from 0.1 to 4.9 µm for Apollo (p > 0.05). Moreover, 90% of the Astra specimens and 70% of the Apollo specimens did not exhibit any microgaps. The highest mean values of microgap size for both groups were detected at the lowest portion of the abutment (p > 0.05). Additionally, the average microgap volume was greater for Apollo than for Astra (p > 0.05). It can be concluded that most samples did not exhibit any microgaps. Furthermore, the linear and volumetric dimensions of microgaps observed at the interface between Apollo or Astra abutments and Astra implants were comparable. Additionally, all tested components presented microgaps (if any) that were clinically acceptable. However, the microgap size of the Apollo abutment was higher and more variable than that of the Astra one.Keywords
Funding Information
- RPLD.01.01.00-10-0002/17-00 (Regional Operational Programme for Łódzkie 2014-2020)
This publication has 75 references indexed in Scilit:
- Implants With Internal Hexagon and Conical Implant-Abutment Connections: An In Vitro Study of the Bacterial ContaminationJournal of Oral Implantology, 2014
- Accuracy combining different brands of implants and abutmentsMedicina Oral Patología Oral y Cirugia Bucal, 2013
- Effects of Different Abutment Connection Designs on the Stress Distribution Around Five Different Implants: A 3-Dimensional Finite Element AnalysisJournal of Oral Implantology, 2012
- Implants with Original and Non‐Original Abutment ConnectionsClinical Implant Dentistry and Related Research, 2012
- Bacterial Leakage in Implants With Different Implant–Abutment Connections: An In Vitro StudyJournal of Periodontology, 2012
- Peri-Implant Crestal Bone Loss: A Putative MechanismInternational Journal of Dentistry, 2012
- Effect of cyclic load on vertical misfit of prefabricated and cast implant single abutmentJournal of Applied Oral Science, 2011
- Sealing Capability and SEM Observation of the Implant-Abutment InterfaceInternational Journal of Dentistry, 2011
- Implant–abutment interface design affects fatigue and fracture strength of implantsClinical Oral Implants Research, 2008
- Role of the Microgap Between Implant and Abutment: A Retrospective Histologic Evaluation in MonkeysJournal of Periodontology, 2003