Detecting electron beam energy shifts with a commercially available energy monitor
Authors:
- Michael D. C. EvansMEMichael D. C. Evans
- Belal A. MoftahBMBelal A. Moftah
- Marina OlivaresMOMarina Olivares
- Ervin B. PodgorsakEPErvin B. Podgorsak
- 11 July 2000
- journal article
- Published by Wiley in Medical Physics
- Vol. 27 (7) , 1661-1663
Abstract
Routine electron beam quality assurance requires an accurate, yet practical, method of energy characterization. Subtle shifts in beam energy may be produced by the linac bending magnet assembly, and the sensitivity of a commercially available electron beam energy‐monitoring device for monitoring these small energy drifts has been evaluated. The device shows an 11% change in signal for a 2 mm change in the energy parameter for low energy electron beams (in the vicinity of 6 MeV) and a 2.5% change in signal for a 2 mm change in the energy parameter for high energy electron beams (in the vicinity of 22 MeV). Thus the device is capable of detecting small energy shifts resulting from bending magnet drift for all clinically relevant electron beams.Keywords
This publication has 3 references indexed in Scilit:
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- Comprehensive QA for radiation oncology: Report of AAPM Radiation Therapy Committee Task Group 40Medical Physics, 1994
- Tests of an electron monitor for routine quality control measurements of electron energiesMedical Physics, 1991