25OHD analogues and vacuum blood collection tubes dramatically affect the accuracy of automated immunoassays
Authors:
Open Access
- 30 September 2015
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 5 (1) , 14636
Abstract
Variations in vitamin D quantification methods are large and influences of vitamin D analogues and blood collection methods have not been systematically examined. We evaluated the effects of vitamin D analogues 25OHD2 and 3-epi 25OHD3 and blood collection methods on vitamin D measurement, using five immunoassay systems and liquid chromatography-tandem mass spectrometry (LC-MS/MS). Serum samples (332) were selected from routine vitamin D assay requests, including samples with or without 25OHD2 or 3-epi 25OHD3 and analysed using various immunoassay systems. In samples with no 25OHD2 or 3-epi 25OHD3, all immunoassays correlated well with LC-MS/MS. However, the Siemens system produced a large positive mean bias of 12.5 ng/mL and a poor Kappa value when using tubes with clot activator and gel separator. When 25OHD2 or 3-epi 25OHD3 was present, correlations and clinical agreement decreased for all immunoassays. Serum 25OHD in VACUETTE tubes with gel and clot activator, as measured by the Siemens system, produced significantly higher values than did samples collected in VACUETTE tubes with no additives. Bias decreased and clinical agreement improved significantly when using tubes with no additives. In conclusion, most automated immunoassays showed acceptable correlation and agreement with LC-MS/MS; however, 25OHD analogues and blood collection tubes dramatically affected accuracy.Keywords
This publication has 26 references indexed in Scilit:
- Vitamin D concentration, obesity, and risk of diabetes: a mendelian randomisation studyThe Lancet Diabetes & Endocrinology, 2014
- Lack of transferability between different immunoassays and LC-MS/MS for total 25-hydroxyvitamin D measurement and disagreement defining deficiencyScandinavian Journal of Clinical and Laboratory Investigation, 2012
- Standardization of measurements of 25-hydroxyvitamin D3 and D2.2012
- Does the presence of 3-epi-25OHD3 affect the routine measurement of vitamin D using liquid chromatography tandem mass spectrometry?Clinical Chemistry and Laboratory Medicine (CCLM), 2012
- 1α,25(OH)2-3-Epi-Vitamin D3, a Natural Physiological Metabolite of Vitamin D3: Its Synthesis, Biological Activity and Crystal Structure with Its ReceptorPLOS ONE, 2011
- Candidate Reference Measurement Procedures for Serum 25-Hydroxyvitamin D3 and 25-Hydroxyvitamin D2 by Using Isotope-Dilution Liquid Chromatography–Tandem Mass SpectrometryClinical Chemistry, 2011
- Preanalytical Stability of 25(OH)–Vitamin D3 in Human Blood or Serum at Room Temperature: Solid as a RockClinical Chemistry, 2009
- Vitamin D DeficiencyThe New England journal of medicine, 2007
- A Prospective Nested Case-Control Study of Vitamin D Status and Pancreatic Cancer Risk in Male SmokersCancer Research, 2006
- C-3 Epimers Can Account for a Significant Proportion of Total Circulating 25-Hydroxyvitamin D in Infants, Complicating Accurate Measurement and Interpretation of Vitamin D StatusThe Journal of Clinical Endocrinology & Metabolism, 2006