Whole-Heart Coronary Magnetic Resonance Angiography
Authors:
- Carsten WarmuthCWCarsten Warmuth
- Jörg SchnorrJSJörg Schnorr
- Nicola KaufelsNKNicola Kaufels
- Susanne WagnerSWSusanne Wagner
- Herbert PilgrimmHPHerbert Pilgrimm
- Bernd HammBHBernd Hamm
- Matthias TaupitzMTMatthias Taupitz
- 1 August 2007
- journal article
- Published by Wolters Kluwer Health in Investigative Radiology
- Vol. 42 (8) , 550-557
Abstract
To test the feasibility and performance of a 4D magnetic resonance coronary angiography sequence compared with conventional inversion recovery (IR) prepared gradient echo imaging. A 4D sequence with 100 milliseconds temporal resolution was implemented on a 1.5 T system. Five minipigs were examined after administration of very small superparamagnetic iron oxide particles. Coronary angiographies with an isotropic resolution of 0.82 mm were performed in the pigs using 4D and IR sequences. The 4D sequence allowed visualization of the coronary arteries, the effect of their movement and that of the entire heart without prolonging scan time. The contrast-to-noise ratio of the IR images was on average 38% higher than that of the corresponding 4D phase. 4D magnetic resonance imaging is superior in that no trigger delay time needs to be determined and an additional whole-heart cine study can be obtained.Keywords
This publication has 24 references indexed in Scilit:
- Coronary Magnetic Resonance Angiography Using Magnetization-Prepared Contrast-Enhanced Breath-Hold Volume-Targeted Imaging (MPCE-VCATS)Investigative Radiology, 2006
- Rest period duration of the coronary arteries: Implications for magnetic resonance coronary angiographyMedical Physics, 2005
- Hochauflösende 3D-MR-Koronarangiographie inEcht-Zeit-Navigatortechnik: Ergebnisse aus107 PatientenuntersuchungenRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2002
- Coronary Magnetic Resonance Angiography for the Detection of Coronary StenosesNew England Journal of Medicine, 2001
- ECG-gated Reconstructed Multi–Detector Row CT Coronary Angiography: Effect of Varying Trigger Delay on Image QualityRadiology, 2001
- 3D magnetization‐prepared true‐FISP: A new technique for imaging coronary arteriesMagnetic Resonance in Medicine, 2001
- Hybrid ordered phase encoding (HOPE): An improved approach for respiratory artifact reductionJournal of Magnetic Resonance Imaging, 1998
- Motion‐adapted gating based on k‐space weighting for reduction of respiratory motion artifactsMagnetic Resonance in Medicine, 1997
- Coronary Angiography with Magnetization‐Prepared T2 ContrastMagnetic Resonance in Medicine, 1995
- Coronary arteries: breath-hold MR angiography.Radiology, 1991