Interference of titanium and zirconia implants on dental-dedicated MR image quality: ex vivo and in vivo assessment
- 18 December 2024
- journal article
- research article
- Published by Oxford University Press (OUP) in Dentomaxillofacial Radiology
- Vol. 54 (2) , 132-139
Abstract
Objectives: To assess the impact of titanium and zirconia implants on dental-dedicated MR image (ddMRI) quality ex vivo (magnetic field distortion [MFD]) and in vivo (artefacts). Methods: ddMR images were acquired (MAGNETOM Free.Max, 0.55 T, Siemens Healthineers AG, Forchheim, Germany) using a dental-dedicated coil (Rapid Biomedical, Rimpar, Germany). Ex vivo: three phantoms were manufactured: one agar-embedded titanium implant, one agar-embedded zirconia implant, and one control phantom (agar 1.5%). Field map analysis of images acquired at 0.55 T, 1.5 T, and 3.0 T (MAGNETOM Sola and MAGNETOM Lumina, respectively, Siemens Healthineers AG, Forchheim, Germany) was done to illustrate the extent and severity of MFD caused by the implants. In vivo (0.55 T only): a splint was designed to serve as an implant carrier, allowing diverse implant positions (0, 1, 2, or 5 implants). A volunteer was imaged using multiple pulse sequences. Three blinded observers scored the images twice for the presence, severity, and type of artefacts, illustrated by descriptive statistics and inter- and intra-observer reproducibility (kappa statistics). Results: Ex vivo: titanium produced more severe MFD than zirconia. MFD extent and amplitude increased with field strength (0.55 T < 1.5 T < 3.0 T). In vivo: titanium produced more artefacts than zirconia, generally as signal voids in tooth crowns close to implants. Inter- and intra-observer reproducibility ranged from 0.28 to 0.64 and 0.32 to 0.57, respectively. Conclusions: The prevalence of artefacts increased with magnetic field strength. Titanium generated larger MFD than zirconia. For both materials, artefacts were visible mainly in the crown area. Observer reproducibility needs improvement by dedicated ddMRI training.Keywords
This publication has 16 references indexed in Scilit:
- Surveying the landscape of diagnostic imaging in dentistry’s futureThe Journal of the American Dental Association, 2024
- Dental-dedicated MRI, a novel approach for dentomaxillofacial diagnostic imaging: technical specifications and feasibilityDentomaxillofacial Radiology, 2023
- The current role and future potential of digital diagnostic imaging in implant dentistry: A scoping reviewClinical Oral Implants Research, 2023
- Magnetic resonance imaging artefacts caused by orthodontic appliances and/or implant-supported prosthesis: a systematic reviewOral Radiology, 2022
- Metallic Implants in MRI – Hazards and Imaging ArtifactsRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2021
- Reliable preparation of agarose phantoms for use in quantitative magnetic resonance elastographyJournal of the Mechanical Behavior of Biomedical Materials, 2019
- Artifact reduction from metallic dental materials in T1‐weighted spin‐echo imaging at 3.0 teslaJournal of Magnetic Resonance Imaging, 2012
- Application of k-space energy spectrum analysis to susceptibility field mapping and distortion correction in gradient-echo EPINeuroImage, 2006
- Dental alloys and magnetic resonance imagingInternational Dental Journal, 2006
- The Kappa Statistic in Reliability Studies: Use, Interpretation, and Sample Size RequirementsPhysical Therapy, 2005