Temperature‐Induced Modulation of the Sample Position in Scanning Electrochemical Microscopy
- 9 April 2015
- journal article
- research article
- Published by Wiley in ChemElectroChem
- Vol. 2 (7) , 946-948
Abstract
Feedback‐mode scanning electrochemical microscopy (SECM) was conducted with in situ control of the sample temperature using a Peltier element. The intrinsic heat pulsing of the temperature control feedback loop causes the sample position to oscillate periodically. By synchronizing tip current acquisition with the oscillatory temperature modulation of the sample, tip current values at different tip‐to‐sample separations are recorded and used to estimate the slope of the distance–current relationship (z‐approach curve). In‐phase amplification of only the distance‐dependent current signal is achieved using the approximated slope and hence the relative imaging contrast is enhanced. Moreover, sample‐position modulated SECM allows distinguishing between electrochemically active and inactive domains on the sample in a single area scan without an a priori knowledge of the sample topography.Keywords
This publication has 13 references indexed in Scilit:
- Mapping Electroactivity at Individual Catalytic Nanostructures Using High-Resolution Scanning Electrochemical–Scanning Ion Conductance MicrocopyAnalytical Chemistry, 2014
- Nanoscale intermittent contact-scanning electrochemical microscopyJournal of Solid State Electrochemistry, 2013
- Structural Correlations in Heterogeneous Electron Transfer at Monolayer and Multilayer Graphene ElectrodesJournal of the American Chemical Society, 2012
- 4D Shearforce-Based Constant-Distance Mode Scanning Electrochemical MicroscopyAnalytical Chemistry, 2010
- Simultaneous Noncontact Topography and Electrochemical Imaging by SECM/SICM Featuring Ion Current Feedback RegulationJournal of the American Chemical Society, 2010
- Intermittent Contact−Scanning Electrochemical Microscopy (IC−SECM): A New Approach for Tip Positioning and Simultaneous Imaging of Interfacial Topography and ActivityAnalytical Chemistry, 2010
- Integrated Ultramicroelectrode−Nanopipet Probe for Concurrent Scanning Electrochemical Microscopy and Scanning Ion Conductance MicroscopyAnalytical Chemistry, 2010
- Constant‐Distance Mode Scanning Electrochemical Microscopy (SECM)—Part I: Adaptation of a Non‐Optical Shear‐Force‐Based Positioning Mode for SECM TipsChemistry – A European Journal, 2003
- Topography feedback mechanism for the scanning electrochemical microscope based on hydrodynamic forces between tip and sampleReview of Scientific Instruments, 1995
- Scanning electrochemical microscopy. 15. Improvements in imaging via tip-position modulation and lock-in detectionAnalytical Chemistry, 1992