Conductive shield for ultra-low-field magnetic resonance imaging: Theory and measurements of eddy currents
- 11 March 2014
- journal article
- Published by AIP Publishing in Journal of Applied Physics
- Vol. 115 (10) , 103902
Abstract
Eddy currents induced by applied magnetic-field pulses have been a common issue in ultra-low-field magnetic resonance imaging. In particular, a relatively large prepolarizing field—applied before each signal acquisition sequence to increase the signal—induces currents in the walls of the surrounding conductive shielded room. The magnetic-field transient generated by the eddy currents may cause severe image distortions and signal loss, especially with the large prepolarizing coils designed for in vivo imaging. We derive a theory of eddy currents in thin conducting structures and enclosures to provide intuitive understanding and efficient computations. We present detailed measurements of the eddy-current patterns and their time evolution in a previous-generation shielded room. The analysis led to the design and construction of a new shielded room with symmetrically placed 1.6-mm-thick aluminum sheets that were weakly coupled electrically. The currents flowing around the entire room were heavily damped, resulting in a decay time constant of about 6 ms for both the measured and computed field transients. The measured eddy-current vector maps were in excellent agreement with predictions based on the theory, suggesting that both the experimental methods and the theory were successful and could be applied to a wide variety of thin conducting structures.Keywords
This publication has 24 references indexed in Scilit:
- Temperature dependence of relaxation times and temperature mapping in ultra-low-field MRIJournal of Magnetic Resonance, 2013
- Hybrid ultra‐low‐field MRI and magnetoencephalography system based on a commercial whole‐head neuromagnetometerMagnetic Resonance in Medicine, 2012
- Evaluation of cancellation coil for precision magnetic measurements with strong prepolarization field inside shielded environmentJournal of Applied Physics, 2012
- Avoiding eddy-current problems in ultra-low-field MRI with self-shielded polarizing coilsJournal of Magnetic Resonance, 2011
- Co-Registration of Interleaved MEG and ULF MRI Using a 7 Channel Low-$T_{\rm c}$ SQUID SystemIEEE Transactions on Applied Superconductivity, 2010
- Pre-polarization enhancement by dynamic nuclear polarization in SQUID-based ultra-low-field nuclear magnetic resonanceSuperconductor Science and Technology, 2010
- Liquid helium cryostat for SQUID-based MRI receiversCryogenics, 2005
- SQUID‐detected MRI at 132 μT with T1‐weighted contrast established at 10 μT–300 mTMagnetic Resonance in Medicine, 2004
- Gradient coil design: A review of methodsMagnetic Resonance Imaging, 1993
- Large-volume conventional magnetic shieldsRevue de Physique Appliquée, 1970