Abstract
Many precision magnetic measurements can benefit significantly from or even require strong prepolarization fields (Bp) and magnetically shielded environments. We describe here in detail a cancellation coil (CC) which can neutralize the Bp on the electrically conductive shield walls that may otherwise induce currents on the walls to produce a lingering transient residual field (Btr) inside the shielded environment and disrupt the measurement operations. The CC was designed using the inverse problem method to effectively neutralize magnetic fields generated on the shield walls by the Bp coil. The implemented CC was evaluated by measuring Btr using a fluxgate magnetometer at different magnetometer positions and cancellation coil currents (ICC). Multi-mode component analysis on the Btr measurements revealed two dominant components, where the component with shorter time constant comes from the current induced around the shield side walls and the other with longer time constant from the current induced on the ceiling and floor of the magnetically shielded room. The analysis also revealed the optimal ICC for each of the top, side, and bottom sections of the CC, which enables significantly easier fine-tuning of individual sections of the CC to enhance CC performance.