Photonics and Optoelectronics are becoming increasingly important for use in implantable devices. In animal trials, although not desirable, failure can lead to the direct replacement of either the component or the test subject. In human implementations, however, the longevity of implantable devices typically needs to exceed 5 years and, in some cases, decades. Traditional hermetic metal packages, per definition, are impervious to water vapour. However, such packaging is unsuitable for structures which are millimetre sized or less. Brain probes encompassing optical micro-emitters must, therefore, use protective passivation/encapsulation layers such as silicon oxynitrides/silicone. However, such protection is prone to electrolytic failure driven by the LED-driving voltages. In this paper, we describe an electrical driving methodology which can improve device lifetime of encapsulated devices by balancing time-averaged electric fields to zero. We have tested the method on commercial optrodes to demonstrate platform independence. We show that, with this method, the time to failure can be increased by over two orders of magnitude.Keywords
Funding Information
- Engineering and Physical Sciences Research Council (NS/A000026/1)
- Wellcome Trust (102037/Z/13/Z)
This publication has 42 references indexed in Scilit:
- Evaluation methods for long-term reliability of polymer-based implantable biomedical devicesBiomedical Engineering Letters, 2021
- A silicon-based neural probe with densely-packed low-impedance titanium nitride microelectrodes for ultrahigh-resolution in vivo recordingsBiosensors and Bioelectronics, 2018
- Opto‐electro‐thermal optimization of photonic probes for optogenetic neural stimulationJournal of Biophotonics, 2018
- Deep Brain Stimulation for Neurological Disorders2015
- A glass-coated tungsten microelectrode enclosing optical fibers for optogenetic exploration in primate deep brain structuresJournal of Neuroscience Methods, 2012
- Visual Prosthetics2011
- Electrical stimulation of excitable tissue: design of efficacious and safe protocolsJournal of Neuroscience Methods, 2005
- Gas sorption, diffusion, and permeation in poly(dimethylsiloxane)Journal of Polymer Science Part B: Polymer Physics, 2000
- Absorption and Scattering of Light by Small Particles1998
- Photo-oxidation of polydimethylsiloxane oils: II—Effect of vinyl groupsPolymer Degradation and Stability, 1992