Entrapment of a J-tip guidewire within the pulmonary vein during FARAPULSETM pulsed field ablation: a case report
Open Access
- 2 July 2026
- journal article
- Published by Oxford University Press (OUP) in European Heart Journal - Case Reports
- Vol. 10 (7)
Abstract
Background: Pulsed field ablation (PFA) is a rapidly adopted technology for atrial fibrillation (AF) ablation and is associated with a favourable safety profile and a low risk of complications. However, mechanical complications have rarely been reported. Case summary: A 37-year-old Asian male with paroxysmal AF underwent PFA using the FARAPULSETM system. Following successful ablation of the left superior pulmonary vein (PV), the J-tip guidewire advanced into the left inferior PV and became completely entrapped. Venography demonstrated that, although the guidewire tip remained intravascular, a proximal segment appeared to course along the vessel wall, suggesting passage through an intraluminal structure. Multiple attempts at withdrawal using push-pull manoeuvres were unsuccessful. Repositioning with a radiofrequency ablation catheter also failed to release the entrapment. Given the risk of vessel rupture and major haemorrhage, the procedure was converted to general anaesthesia in a hybrid operating room with surgical standby. Bridge occlusion balloon-assisted extraction successfully released the entrapped guidewire without vascular injury. The retrieved guidewire had adherent fibrous white tissue, suggesting entrapment by a membranous pulmonary venous structure. Discussion: Balloon-assisted endovascular retrieval may represent a safe and less invasive alternative to surgical intervention for managing such entrapments. Operators should be aware of this rare but potentially hazardous complication.Keywords
This publication has 16 references indexed in Scilit:
- Guidewire entrapment in pulmonary veins during pulsed field ablation of atrial fibrillation: A case seriesHeartRhythm Case Reports, 2026
- Role of the Bridge balloon in improving the safety of transvenous lead extraction proceduresHeart Rhythm, 2025
- Guidewire entrapment during pulsed field ablation: A case reportHeartRhythm Case Reports, 2025
- Entrapment of the PulseSelect™ catheter in the pulmonary vein: A rare complication due to soft tissue obstructionJournal of Arrhythmia, 2025
- Pulsed Field or Cryoballoon Ablation for Paroxysmal Atrial FibrillationNew England Journal of Medicine, 2025
- 2024 ESC Guidelines for the management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS)European Heart Journal, 2024
- Safety of pulsed field ablation in more than 17,000 patients with atrial fibrillation in the MANIFEST-17K studyNature Medicine, 2024
- Entrapment of ablation catheter in right pulmonary veins requiring surgery to removeHeartRhythm Case Reports, 2024
- Pulsed Field Ablation for Pulmonary Vein Isolation in Atrial FibrillationJACC, 2019
- A case of an ablation catheter entrapped in the pulmonary vein during atrial fibrillation ablation requiring open heart surgery for removalHeartRhythm Case Reports, 2017