Estimation of the focal spot size and shape for a medical linear accelerator by Monte Carlo simulation
Authors:
- Lilie L. W. WangLWLilie L. W. Wang
- Konrad LeszczynskiKLKonrad Leszczynski
- 17 January 2007
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 34 (2) , 485-488
Abstract
The focal spot size and shape of a medical linac are important parameters that determine the dose profiles, especially in the penumbral region. A relationship between the focal spot size and the dose profile penumbra has been studied and established from simulation results of the EGSnrc Monte Carlo code. A simple method is proposed to estimate the size and the shape of a linac's focal spot from the measured dose profile data.Keywords
This publication has 17 references indexed in Scilit:
- Geometrical and dosimetrical characterization of the photon source using a micro-multileaf collimator for stereotactic radiosurgeryPhysics in Medicine & Biology, 2003
- Sensitivity of megavoltage photon beam Monte Carlo simulations to electron beam and other parametersMedical Physics, 2002
- Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 versionMedical Physics, 2000
- Performance and beam characteristics of the Siemens Primus linear acceleratorMedical Physics, 1999
- X‐ray source and the output factorMedical Physics, 1995
- BEAM: A Monte Carlo code to simulate radiotherapy treatment unitsMedical Physics, 1995
- The source-density function: determination from measured lateral dose distributions and use for convolution dosimetryPhysics in Medicine & Biology, 1993
- Dosimetry of 6-MV x-ray beam penumbraMedical Physics, 1993
- X‐ray sources of medical linear accelerators: Focal and extra‐focal radiationMedical Physics, 1993
- Therapy imaging: Source sizes of radiotherapy beamsMedical Physics, 1988