Ability of Two Commercially Available Assays (Abbott RealTi m e HIV-1 and Roche Cobas AmpliPrep/Cobas TaqMan HIV-1 Version 2.0) To Quantify Low HIV-1 RNA Levels (<1,000 Copies/Milliliter): Comparison with Clinical Samples and NIBSC Working Reagent for Nucleic Acid Testing Assays
- 1 June 2014
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 52 (6) , 2019-2026
Abstract
Concordance between molecular assays may be suboptimal at low HIV-1 viremia levels (m e HIV-1 and the Roche Cobas AmpliPrep/TaqMan HIV-1 version 2.0) for their ability to detect and quantify low viral loads. A comparison was performed using 167 residual clinical samples (with values ranging from “not detected” to 1,000 copies/ml, as measured by the Abbott assay) and the National Institute and Biological Standards and Control (NIBSC) HIV-1 RNA working reagent 1 for nucleic acid amplification techniques (NAT) assays (serially diluted to a range from 1 to 1,000 copies/ml). Quantitative results were compared using Lin's concordance correlation coefficient and a Bland-Altman plot. Concordance with the qualitative results was measured by Cohen's kappa statistic. With clinical samples, the degree of interassay concordance of the qualitative results at a 40-copies/ml HIV-1 RNA threshold was substantial (κ = 0.762); the correlation among the quantified samples was suboptimal (concordance correlation coefficient, 0.728; P < 0.0001); the mean difference of the values between the Roche and Abbott assays was 0.193 log 10 copies/ml. Using the HIV-1 RNA working reagent 1 for NAT assays, the results provided by the Roche assay were, on average, 3 times higher than expected, while the Abbott assay showed high accuracy. The Roche assay was highly sensitive, being able to detect a level as low as 3.5 copies/ml HIV-1 RNA with 95% probability. The performance characteristics of each molecular assay should be taken into account when HIV-1 RNA threshold values for “virologic suppression,” “virologic failure,” “persistent low viral loads,” etc., are defined and indicated in the support of clinical decisions.Keywords
This publication has 40 references indexed in Scilit:
- Cobas Ampliprep/Cobas TaqMan HIV-1 v2.0 Assay: Consequences at the Cohort LevelPLOS ONE, 2013
- Quantification of Viral Loads Lower than 50 Copies per Milliliter by Use of the Cobas AmpliPrep/Cobas TaqMan HIV-1 Test, Version 2.0, Can Predict the Likelihood of Subsequent Virological Rebound to >50 Copies per MilliliterJournal of Clinical Microbiology, 2013
- Increased Risk of Virologic Rebound in Patients on Antiviral Therapy with a Detectable HIV Load <48 Copies/mLPLOS ONE, 2012
- Impact of Low-Level-Viremia on HIV-1 Drug-Resistance Evolution among Antiretroviral Treated-PatientsPLOS ONE, 2012
- HIV-1 low-level viraemia assessed with 3 commercial real-time PCR assays show high variabilityBMC Infectious Diseases, 2012
- Magnitude of Virologic Blips Is Associated With a Higher Risk for Virologic Rebound in HIV-Infected Individuals: A Recurrent Events AnalysisThe Journal of Infectious Diseases, 2012
- Antiretroviral Drug Resistance in HIV-1–Infected Patients Experiencing Persistent Low-Level Viremia During First-Line TherapyThe Journal of Infectious Diseases, 2011
- Comparative Evaluation of the Roche Cobas AmpliPrep/Cobas TaqMan HIV-1 Version 2 Test Using the TaqMan 48 Analyzer and the Abbott RealTi m e HIV-1 AssayJournal of Clinical Microbiology, 2011
- Impact of Discrepancies between the Abbott RealTime and Cobas TaqMan Assays for Quantification of Human Immunodeficiency Virus Type 1 Group M Non-B SubtypesJournal of Clinical Microbiology, 2009
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986