High-Frame-Rate Doppler Ultrasound Using a Repeated Transmit Sequence
Open Access
- 1 February 2018
- journal article
- research article
- Published by MDPI AG in Applied Sciences
- Vol. 8 (2) , 227
Abstract
The maximum detectable velocity of high-frame-rate color flow Doppler ultrasound is limited by the imaging frame rate when using coherent compounding techniques. Traditionally, high quality ultrasonic images are produced at a high frame rate via coherent compounding of steered plane wave reconstructions. However, this compounding operation results in an effective downsampling of the slow-time signal, thereby artificially reducing the frame rate. To alleviate this effect, a new transmit sequence is introduced where each transmit angle is repeated in succession. This transmit sequence allows for direct comparison between low resolution, pre-compounded frames at a short time interval in ways that are resistent to sidelobe motion. Use of this transmit sequence increases the maximum detectable velocity by a scale factor of the transmit sequence length. The performance of this new transmit sequence was evaluated using a rotating cylindrical phantom and compared with traditional methods using a 15-MHz linear array transducer. Axial velocity estimates were recorded for a range of mm/s and compared to the known ground truth. Using these new techniques, the root mean square error was reduced from over 400 mm/s to below 50 mm/s in the high-velocity regime compared to traditional techniques. The standard deviation of the velocity estimate in the same velocity range was reduced from 250 mm/s to 30 mm/s. This result demonstrates the viability of the repeated transmit sequence methods in detecting and quantifying high-velocity flow.
Keywords
This publication has 14 references indexed in Scilit:
- High-Frame-Rate Echocardiography Using Coherent Compounding With Doppler-Based Motion-CompensationIEEE Transactions on Medical Imaging, 2016
- Coherent compounding in doppler imagingIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2015
- Coherent Plane Wave Compounding for Very High Frame Rate Ultrasonography of Rapidly Moving TargetsIEEE Transactions on Medical Imaging, 2013
- Comparison of conventional parallel beamforming with plane wave and diverging wave imaging for cardiac applications: a simulation studyIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2012
- GPU-based beamformer: Fast realization of plane wave compounding and synthetic aperture imagingIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2011
- High-frame-rate echocardiography using diverging transmit beams and parallel receive beamformingJournal of Medical Ultrasonics, 2011
- Colour flow and motion imagingProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2009
- Coherent plane-wave compounding for very high frame rate ultrasonography and transient elastographyIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2009
- Extended high-frame rate imaging method with limited-diffraction beamsIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2006
- 2D and 3D high frame rate imaging with limited diffraction beamsIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1997