Effect of protocol choice on phase contrast cardiac magnetic resonance flow measurement in the ascending aorta: breath-hold and non-breath-hold
- 22 April 2012
- journal article
- research article
- Published by Springer Nature in The International Journal of Cardiovascular Imaging
- Vol. 29 (1) , 113-120
Abstract
Flow assessment with phase contrast magnetic resonance imaging (PC-MRI) protocols is an important component of a comprehensive cardiovascular MR (CMR) assessment. Breath-hold (BH) and non-breath-hold (NBH) PC-MRI protocols are widely available for this imaging modality. Because flow in the great vessels is known to vary with the respiratory cycle, we hypothesized that these 2 approaches might yield different results in the clinical assessment of forward and regurgitant flow in the ascending aorta. Further, given renewed awareness of the possible effect of velocity offsets in PC-MRI, we also sought to evaluate the impact of BH and NBH protocols on this potential source of error. A prospective observational study was performed in 55 consecutive patients referred for clinical CMR of the thoracic aorta. Both BH and NBH protocols were performed at the sinotubular junction and at the mid ascending aorta. Ten additional patients underwent repeated scanning at the mid ascending aorta with both BH and NBH protocols so that protocol variability could be assessed. Finally, ten patients were scanned with both BH and NBH protocols, and phantoms were then imaged with identical imaging parameters so that offset errors associated with each protocol could be evaluated. Forward flow was generally greater with the NBH protocol than with the BH protocol (mean values 102.1 mL vs. 97.9 mL; P = 0.0004). The Bland–Altman limits of agreement were quite wide for all indices (e.g, forward flow, −26.7 mL, +18.2 mL), which suggests that results from BH and NBH protocols cannot be interchanged with confidence. Estimated phase offset errors were similar for both protocols and were generally within acceptable ranges at the mid ascending level, with slightly higher values observed at the sinotubular junction for the BH technique. We observed differences in flow values with BH and NBH protocols for PC-MRI. This finding is relevant to patients imaged serially for the evaluation of cardiac output or valve (aortic or mitral) insufficiency, for whom adherence to one PC-MRI breathing protocol is likely most effective.Keywords
This publication has 20 references indexed in Scilit:
- Aortic dissection: visualisation of aortic blood flow and quantification of wall shear stress using time-resolved, 3D phase-contrast MRIJournal of Cardiovascular Magnetic Resonance, 2011
- Sequence optimization to reduce velocity offsets in cardiovascular magnetic resonance volume flow quantification - A multi-vendor studyJournal of Cardiovascular Magnetic Resonance, 2011
- Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurementsJournal of Cardiovascular Magnetic Resonance, 2010
- The effects of breath-holding on pulmonary regurgitation measured by cardiovascular magnetic resonance velocity mappingJournal of Cardiovascular Magnetic Resonance, 2009
- Baseline Correction of Phase Contrast Images Improves Quantification of Blood Flow in the Great VesselsJournal of Cardiovascular Magnetic Resonance, 2007
- Flow Measurement by Magnetic Resonance: A Unique Asset Worth OptimisingJournal of Cardiovascular Magnetic Resonance, 2007
- Cardiorespiratory‐resolved magnetic resonance imaging: Measuring respiratory modulation of cardiac functionMagnetic Resonance in Medicine, 2006
- Clinical blood flow quantification with segmented k‐space magnetic resonance phase velocity mappingJournal of Magnetic Resonance Imaging, 2002
- The application of phase shifts in NMR for flow measurementMagnetic Resonance in Medicine, 1990
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986