Abstract
As microfluidics has been applied extensively in many cell and biochemical applications, monitoring the related processes is an important requirement. In this work, we design and fabricate a high-throughput microfluidic device which contains 32 microchambers to perform automated parallel microfluidic operations and monitoring on an automated stage of a microscope. Images are captured at multiple spots on the device during the operations for monitoring samples in microchambers in parallel; yet the device positions may vary at different time points throughout operations as the device moves back and forth on a motorized microscopic stage. Here, we report an image-based positioning strategy to realign the chamber position before every recording of microscopic image. We fabricate alignment marks at defined locations next to the chambers in the microfluidic device as reference positions. We also develop image processing algorithms to recognize the chamber positions in real-time, followed by realigning the chambers to their preset positions in the captured images. We perform experiments to validate and characterize the device functionality and the automated realignment operation. Together, this microfluidic realignment strategy can be a platform technology to achieve precise positioning of multiple chambers for general microfluidic applications requiring long-term parallel monitoring of cell and biochemical activities.Keywords
Funding Information
- City University of Hong Kong (7200267, 9610212, RGC115813)
- City University of Hong Kong (7200267, 9610212, RGC115813)
- General Research Fund of Hong Kong Research Grant Council (7200267, 9610212, RGC115813)
- Croucher Foundation (7200267, 9610212, RGC115813)
This publication has 36 references indexed in Scilit:
- Real-time and non-invasive impedimetric monitoring of cell proliferation and chemosensitivity in a perfusion 3D cell culture microfluidic chipBiosensors and Bioelectronics, 2014
- High throughput capture of circulating tumor cells using an integrated microfluidic systemBiosensors and Bioelectronics, 2013
- Digital microfluidics with impedance sensing for integrated cell culture andanalysisBiosensors and Bioelectronics, 2013
- Long-term three-dimensional cell culture and anticancer drug activity evaluation in a microfluidic chipBiosensors and Bioelectronics, 2013
- Frequency discretization in dielectrophoretic assisted cell sorting arrays to isolate neural cellsLab on a Chip, 2012
- Microfluidics for Mammalian Cell ChemotaxisAnnals of Biomedical Engineering, 2011
- Microfluidic sorting of mammalian cells by optical force switchingNature Biotechnology, 2005
- Continuous perfusion microfluidic cell culture array for high‐throughput cell‐based assaysBiotechnology and Bioengineering, 2004
- An Integrated Microfluidic System for Reaction, High-Sensitivity Detection, and Sorting of Fluorescent Cells and ParticlesAnalytical Chemistry, 2003
- Development of a microfluidic device for fluorescence activated cell sortingJournal of Micromechanics and Microengineering, 2002