Dual Chamber Pacemaker Implant in Coronary Sinus Leading to Several Complications
Open Access
- 5 November 2024
- journal article
- research article
- Published by MDPI AG in Diagnostics
- Vol. 14 (22) , 2465
Abstract
Permanent pacemaker implantation is a low-risk procedure. However, complications may occur at a rate of around 4–8%. We present a case where initial implantation resulted in complications that could have been avoided by meticulous assessment of lead position in different projections and early post-procedure X-ray that would have delineated other serious complications. We present a case where the right ventricular lead was placed in the coronary sinus, which resulted in the loss of pacing capture with further syncope after the pacemaker implant. This was apparent in the post-procedure electrocardiogram (ECG) with right bundle branch pacing and the lead was repositioned in the right ventricular apex the following day. Furthermore, the patient was discharged home without a chest X-ray (CXR), and she represented a week later with a haemo-pneumothorax and pericardial effusion. A chest drain was placed and was discharged after a slow recovery following several complications that could have been avoidable.Keywords
All Related Versions
This publication has 6 references indexed in Scilit:
- Pacemaker Interrogation and Programming2022
- Chest radiographs of cardiac devices (Part 1): Cardiovascular implantable electronic devices, cardiac valve prostheses and Amplatzer occluder devices (vol 23, a1730, 2019)South African Journal of Radiology, 2021
- Chest radiographs of cardiac devices (Part 1): Cardiovascular implantable electronic devices, cardiac valve prostheses and Amplatzer occluder devicesSouth African Journal of Radiology, 2019
- Safety of Permanent Pacemaker Implantation: A Prospective StudyJournal of Clinical Medicine, 2019
- Individualized Left Anterior Oblique ProjectionCirculation: Arrhythmia and Electrophysiology, 2018
- A case of difficult RV lead placementHeart, 2013