Transient adhesion of platelets in pump-oxygenator systems: influence of SMA and nitric oxide treatments

Authors:

Abstract
We employed gamma scintigraphy to quantify the transient accumulations of platelets in pump-oxygenator systems employed in cardiopulmonary bypass (CPB). A flat sheet microporous polypropylene membrane oxygenator (Cobe Duo) was employed, with and without siloxane/caprolactone oligomer coating (SMA) (n = 8 each). The effect of nitric oxide gas infusion on platelet deposition was also evaluated for the uncoated Cobe Duo system (n = 10 each). Scintigraphic images of radiolabelled cells were obtained and converted to numbers of all platelets, labeled and unlabeled, adhering to the pump and oxygenator surfaces. These numbers were compared, by study group, for a 90-min period of normothermic CPB in the adult pig, employing standard prime and anticoagulation regimens. Platelets adhered in large numbers to control oxygenators, reaching maxima (> 20% of the circulating platelet mass) 30 min following institution of CPB, and decreasing for the duration of CPB. SMA treatment significantly decreased platelet adhesion ...