Abstract
To determine whether monomorphic ventricular tachycardias (VTs) have different characteristics and/or responses to antitachycardia pacing (ATP) with respect to the indication—primary prevention (PP) versus secondary prevention (SP)—among ICD patients with left ventricular dysfunction. We prospectively studied 551 VT (cycle length [CL] 329 ± 35 ms; PP 34 %) occurring in 67 ICD patients with left ventricular dysfunction (LVEF 35 ± 8 %). ICD programming was standardized, including ATP for slow (CL 400–321 ms) and fast VT (CL 250–320 ms). We analyzed the following aspects: CL, percentage of variability of the 12 RR intervals prior to ATP (P-RR)—which was calculated by dividing the mean difference between each R-R interval with the next one by the CL × 100—and type of termination: immediate (VT ceased immediately upon ATP completion) or delayed (VT persisted after ATP). ATP was successful in 86 % of VTs. VTs occurring in SP patients had a lower P-RR, median (IQR) 2.7 % (1.2–3.7) versus 1.9 % (0.9–3.2); p = 0.002; they terminated immediately after ATP less frequently (27 % versus 12 %; p < 0.001), and although they were more frequently slow (51 % versus 67 %; p = 0.01), ATP was less effective in them, 92 versus 80 % (p = 0.02). VTs occurring in SP patients are slower, more stable, and they terminate less frequently at ATP. Therefore, compared with PP, SP patients seem to have fewer self-terminating VTs.

This publication has 21 references indexed in Scilit: