Measurement uncertainty in clinical chemistry: ISO 20914 versus nordtest or intermediate precision versus bias
- 2 May 2024
- journal article
- research article
- Published by Taylor & Francis in Scandinavian Journal of Clinical and Laboratory Investigation
- Vol. 84 (3) , 147-153
Abstract
Measuring uncertainty (MU) is crucial to ensure the accuracy and precision of laboratory results. This study compares the ISO 20914 and Nordtest guidelines to analyze the MU values for 20 clinical chemistry analytes over six months. The researchers calculated MU components, including within-laboratory reproducibility (Rw), laboratory analytical performance bias (u(bias)), and combined standard uncertainty (uc), based on internal quality control and external quality assessment data. The final expanded uncertainty (U) values were determined by multiplying the combined uncertainty with a coverage factor (k = 2 for 95% Confidence Interval), following each guideline's respective procedures. Clinical chemistry analytes were analyzed on Roche Cobas 6000 c501 auto analyzer (Roche Diagnostics, Mannheim, Germany) and manufacturer's kits were used analysis. The results show that 11 out of 20 clinical chemistry analytes met the targeted maximum allowable measurement uncertainty (MAU) values when calculated according to ISO 20914 guideline. Also, 11 out of 20 clinical chemistry analytes' MU values met the MAU values with the Nordtest guideline's recommended calculations. However, some tests met the MAU in the ISO 20914 approach but not in the Nordtest guideline, and vice versa. The study found that intermediate precision (uRw) in the ISO 20914 approach and performance bias (u(bias)) in the Nordtest approach significantly impacted MU values. The research highlights the importance of standardization in MU calculation approaches across clinical laboratories. These findings have implications for patient care and clinical decision-making, emphasizing the importance of selecting appropriate laboratory guidelines for routine use.Keywords
This publication has 24 references indexed in Scilit:
- Bias, the unfinished symphonyBiochemia Medica, 2022
- Sigma metrics in laboratory medicine: A call for harmonizationClinica Chimica Acta, 2022
- Measurement uncertainty for practical useClinica Chimica Acta, 2022
- Interpreting EQA—Understanding Why Commutability of Materials MattersClinical Chemistry, 2022
- The top-down approach to measurement uncertaintyBiochemia Medica, 2020
- Deriving proper measurement uncertainty from Internal Quality Control data: An impossible mission?Clinical Biochemistry, 2018
- What information on measurement uncertainty should be communicated to clinicians, and how?Clinical Biochemistry, 2018
- Minimum requirements for the estimation of measurement uncertainty: Recommendations of the joint Working group for uncertainty of measurement of the CSMBLM and CCMBBiochemia Medica, 2017
- Pre-analytical and analytical aspects affecting clinical reliability of plasma glucose resultsClinical Biochemistry, 2017
- Update on current concepts and meanings in laboratory medicine — Standardization, traceability and harmonizationClinica Chimica Acta, 2014