Measurement of temporal resolution in dual source CT
- 29 January 2008
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 35 (2) , 764-768
Abstract
This work assesses the temporal resolution of dual‐source computed tomography (CT) in a visually intuitive manner. Exploiting the principles of ring artifact creation, a phantom containing a highly attenuating delta function signal (a wire) was rotated at the same speed as the gantry, creating a partial ring artifact where the angular extent of the artifact provides a direct indication of the temporal resolution. A demonstration of the effect of the evaluated 165 and nominal temporal resolutions on motion artifacts in cardiac CT is provided using patient data.Keywords
This publication has 8 references indexed in Scilit:
- Diagnostic accuracy of dual-source multi-slice CT-coronary angiography in patients with an intermediate pretest likelihood for coronary artery diseaseEuropean Heart Journal, 2007
- Evaluation of Coronary Stents and Stenoses at Different Heart Rates With Dual Source Spiral CT (DSCT)Investigative Radiology, 2007
- Dose Performance of a 64-Channel Dual-Source CT Scanner1Radiology, 2007
- Dual-Source Computed TomographyInvestigative Radiology, 2007
- Dual-source CT cardiac imaging: initial experienceEuropean Radiology, 2006
- First performance evaluation of a dual-source CT (DSCT) systemEuropean Radiology, 2005
- Image reconstruction and image quality evaluation for a 64‐slice CT scanner with ‐flying focal spotMedical Physics, 2005
- Heart Rate Adaptive Optimization of Spatial and Temporal Resolution for Electrocardiogram-Gated Multislice Spiral CT of the HeartJournal of Computer Assisted Tomography, 2001