Practical time considerations for optically stimulated luminescent dosimetry (OSLD) in total body irradiation
- 13 February 2017
- journal article
- scientific note
- Published by Springer Nature in Australasian Physics & Engineering Sciences in Medicine
- Vol. 40 (1) , 167-171
Abstract
Total body irradiation (TBI) treatments are used to treat the whole body in preparation for hematopoietic stem cell (or bone marrow) transplantation. Our standard clinical regimen is a 12 Gy in 6 fraction, bi-daily technique using 6 MV X-rays at an extended Source-to-Surface distance (SSD) of 300 cm. Utilizing these characteristics, the beam dose rate is reduced below 7 cGy/min as is standard for TBI treatment. Dose received by the patient is monitored using optically stimulated luminescent dosimetry (OSLD). This work presents some practical calibration corrections based on time-dependant factors for OSLD calibration related to TBI procedure. Results have shown that a negligible difference is seen in OSL sensitivity for 6 MV X-rays irradiated in standard SSD (100 cm) and high dose rate (600 cGy/min) conditions compared to extended SSD (300 cm) and low TBI dose rate (6 cGy/min) conditions. Results have also shown that whilst short term signal fading occurs in the OSL after irradiation at a high dose rate (37% reduction in signal in the first 15 min), thereafter, negligible differences are seen in the OSL signal between 600 and 7 cGy/min irradiations. Thus a direct comparison can be made between calibration OSLs and clinical TBI OSLs between 15 min and 2 h. Finally a table is presented to provide corrections between calibration OSL readout and clinical TBI dose readout for a period up to 7 days. Combining these three results allows users to pre-irradiate their calibration OSLs at standard dose rate and SSD, up to 1 week prior to clinical treatment, and still provide accurate in-vivo dosimetry. This can help with time saving and work efficiency in the clinic.Keywords
This publication has 10 references indexed in Scilit:
- SU-E-T-600: In Vivo Dosimetry for Total Body and Total Marrow Irradiations with Optically Stimulated Luminescence DosimetersMedical Physics, 2014
- Benefits of online in vivo dosimetry for single-fraction total body irradiationMedical Dosimetry, 2014
- SU-E-T-85: Total Body Irradiation (TBI) Optically Stimulated Luminescence In Vivo DosimetryMedical Physics, 2013
- Commissioning of optically stimulated luminescence dosimeters for use in radiotherapyRadiation Measurements, 2013
- Clinical Experience in the use of Optically Stimulated Luminescent Dosimeters for Total Body IrradiationInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- In-vivo dosimetry from total body irradiation patients (2000–2006): results and analysisAustralasian Physics & Engineering Sciences in Medicine, 2008
- Patient dosimetry for total body irradiation using single‐use MOSFET detectorsJournal of Applied Clinical Medical Physics, 2008
- Clinical application of GAFCHROMIC® EBT film for in vivo dose measurements of total body irradiation radiotherapyApplied Radiation and Isotopes, 2008
- Dosimetry and verification of 60Co total body irradiation with human phantom and semiconductor diodesJournal of Medical Physics, 2007
- In-vivo dosimetry by diode semiconductors in combination with portal films during TBI: reporting a 5-year clinical experienceRadiotherapy and Oncology, 1999