Minimum Lyophilized Plasma Requirement for ISI Calibration
Open Access
- 1 February 1998
- journal article
- research article
- Published by Oxford University Press (OUP) in American Journal of Clinical Pathology
- Vol. 109 (2) , 196-204
Abstract
The minimum requirement of lyophilized plasma samples for a reliable International Sensitivity Index (ISI) was assessed by calibrations based on reducing numbers from a maximum of 60 depleted and 20 plasma samples from patients receiving coumarin using a manual technique and low ISI thromboplastin. The probability of achieving an international normalized ratio (INR) result within a clinically important range of 20% deviation has been assessed at European Concerted Action on Anticoagulation (ECAA) national laboratories in calibrations of ECAA reference thromboplastin against the certified prothrombin time (FT) with a reference thromboplastin. With conventional orthogonal regression using a certified PT and linear regression using certified INR, deviations and the coefficient of variation of the calibration slope increased with reduced numbers. The INR deviations became marked when the number of abnormal plasma samples was reduced to fewer than 20. Calibrations of 3 abnormal plasma samples and 1 lyophilized normal plasma sample gave a high incidence of deviations greater than 10% with an INR of 3.0. The study demonstrates that with both methods of analysis, an optimum minimum number of lyophilized plasma samples is needed for a reliable local ISI.Keywords
This publication has 6 references indexed in Scilit:
- A simplified statistical method for local INR using linear regressionBritish Journal of Haematology, 1997
- A Simplified Procedure for Thromboplastin Calibration -The Usefulness of Lyophilized Plasmas Assessed in a Collaborative StudyThrombosis and Haemostasis, 1996
- European Concerted Action on Anticoagulation (ECAA) - The Multicentre Calibration of Rabbit and Human ECAA Reference ThromboplastinsThrombosis and Haemostasis, 1996
- A Comparison of Lyophilized Artificially Depleted Plasmas and Lyophilized Plasmas From Patients Receiving Warfarin in Correcting for Coagulometer Effects on International Normalized RatiosAmerican Journal of Clinical Pathology, 1995
- The Value of Plasma Calibrants in Correcting Coagulometer Effects on International Normalized Ratios:An International Multicenter StudyAmerican Journal of Clinical Pathology, 1995
- Assessment of value of calibrated lyophilised plasmas to determine International Sensitivity Index for coagulometers.Journal of Clinical Pathology, 1992