Reuse of Infected Cardiac Rhythm Management Devices in the Same Patients: A Single-Center Experience
- 5 February 2014
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 37 (8) , 940-946
Abstract
In developing countries, patients with a cardiac rhythm management device (CRMD) that has become infected cannot always afford a new device, and reuse of the infected CRMD may be appropriate. To evaluate the safety of this practice, a cohort of consecutive patients with reused pacemakers was compared with a control group. A single-center cohort of consecutive patients treated from 2007 to 2012 was analyzed in a two-way noninferiority study. Patients who had infected CRMDs removed at the Cardiovascular Center of The Peking University People's Hospital were enrolled, including those who had the same CRMD reimplanted and those who had a new CRMD implanted. This study included 212 patients, of who 99 (study group) had the same CRMD reimplanted, and 113 (control group) had a new CRMD implanted. During a mean follow-up period of 3.52 years (interquartile range: 2.1-4.7 years), 10 patients reached the primary end point of the study (infection, unexpected battery depletion, or device malfunction), including five (5%) in the study group and five (4.3%) in the control group. Recurrent infection occurred in three patients in the study group and two patients in the control group (3.0% vs 1.7%, relative risk: 1.29, 95% confidence interval: 0.62-2.29, P = 0.561). There were no cases of unexpected battery depletion in either group. Reuse of infected CRMDs in the same patients is noninferior to the implantation of new devices. In developing countries without full medical coverage, careful reuse of infected CRMDs can be considered.Keywords
Funding Information
- National Key Basic Research program (NKBRP) of China (2007CB512008)
- Key Program of the Beijing Science and Technology Commission (D111100003011003, D111100003011006, Z121107001012016)
This publication has 20 references indexed in Scilit:
- The ethics of pacemaker reuse: might the best be the enemy of the good?Heart, 2011
- Safety of Pacemaker ReuseCirculation: Arrhythmia and Electrophysiology, 2011
- Safety, Efficacy, and Performance of Implanted Recycled Cardiac Rhythm Management (CRM) Devices in Underprivileged PatientsPacing and Clinical Electrophysiology, 2011
- Cardiac Device Reutilization: Is It Time to “Go Green” in Underserved Countries?Pacing and Clinical Electrophysiology, 2011
- Pacemaker ReuseCirculation, 2010
- The World Survey of Cardiac Pacingand Cardioverter‐Defibrillators: Calendar Year 2005 An International Cardiac Pacing andElectrophysiology Society (ICPES) projectPacing and Clinical Electrophysiology, 2008
- Rising Rates of Cardiac Rhythm Management Device Infections in the United States: 1996 through 2003JACC, 2006
- Re-used pacemakers—as safe as new? A retrospective case-control studyEuropean Heart Journal, 1998
- Report of the Policy Conference on Pacemaker Re-use Sponsored by the North American Society of Pacing and ElectrophysiologyPacing and Clinical Electrophysiology, 1985
- The Refurbished Puise GeneratorPacing and Clinical Electrophysiology, 1980