In vitrodissolution of plasma-sprayed hydroxyapatite coatings with different characteristics: experimental study and modeling
Authors:
- Z MohammadiZMZ Mohammadi
- A A Ziaei-MoayyedAZA A Ziaei-Moayyed
- A Sheikh-Mehdi MesgarAMA Sheikh-Mehdi Mesgar
- 8 January 2008
- journal article
- research article
- Published by IOP Publishing in Biomedical Materials
- Vol. 3 (1) , 015006
Abstract
The dissolution of plasma-sprayed hydroxyapatite (PHA) coatings with different characteristics, produced by various spraying conditions, in a Tris-buffered solution at pH 7.4 was experimentally studied through the measurement of the release of calcium ions. The phase composition of the coatings at surface and interface, and the porosity were evaluated. The analytical modeling revealed that the calcium dissolution process was composed of two stages. The first stage was found to be both surface and diffusion controlled. The second stage was an exactly diffusion-controlled dissolution. In the first stage, the rate of dissolution and the solubility of the coatings with minimum contents of impurity phases were mainly influenced by the contents of recrystallized HA (RHA) and amorphous calcium phosphate (ACP). It is suggested that the optimized values of the ACP and the RHA at the coating surface can tend to encourage the early fixation properties of the PHA coatings.Keywords
This publication has 40 references indexed in Scilit:
- Grit blasting of Ti–6Al–4V alloy: Optimization and its effect on adhesion strength of plasma-sprayed hydroxyapatite coatingsJournal of Materials Processing Technology, 2007
- A review on calcium phosphate coatings produced using a sputtering process—an alternative to plasma sprayingBiomaterials, 2005
- Thermal sprayed hydroxyapatite splats: nanostructures, pore formation mechanisms and TEM characterizationBiomaterials, 2004
- Adsorption/Desorption Phenomena of Plasma Sprayed Hydroxyapatite Coating Surface during Immersion in the Ca-Free SBFJournal of the Ceramic Society of Japan, 2004
- Dissolution and mineralization behaviors of HA coatingsBiomaterials, 2003
- Surface characteristics and dissolution behavior of plasma‐sprayed hydroxyapatite coatingJournal of Biomedical Materials Research, 2002
- Mechanical and chemical consequences of the residual stresses in plasma sprayed hydroxyapatite coatingsBiomaterials, 1997
- Degradation potential of plasma‐sprayed hydroxyapatite‐coated titanium implantsJournal of Biomedical Materials Research, 1995
- Biodegradation and bioresorption of calcium phosphate ceramicsClinical Materials, 1993
- In vitro evaluation of amorphous calcium phosphate and poorly crystallized hydroxyapatite coatings on titanium implantsJournal of Biomedical Materials Research, 1993