Transcutaneous carbon dioxide monitoring in infants and children
- 24 April 2009
- journal article
- review article
- Published by Wiley in Pediatric Anesthesia
- Vol. 19 (5) , 434-444
Abstract
Summary: Objective: To review the technology required for and the applications of transcutaneous carbon dioxide (TC‐CO2) monitoring in infants and children.Data source: A computerized, bibliographic search regarding the applications of transcutaneous carbon dioxide (TC‐CO2) monitoring in infants and children.Results: Although the direct measurement of PaCO2 remains the gold standard, it provides only a single measurement of what is often a rapidly changing and evolving clinical picture. Given these concerns, there remains a clinical need for a means to continuously monitor PaCO2 without the need for repeated blood gas analysis. Although initially introduced into the neonatal intensive care unit; with improvements in the technology, TC‐CO2 monitoring can now be used in infants, children and even adults. When compared with end‐tidal carbon dioxide (ET‐CO2) monitoring techniques, TC‐CO2 monitoring has been shown to be equally as accurate in patients with normal respiratory function and more accurate in patients with shunt or ventilation–perfusion inequalities. TC‐CO2 monitoring can be applied in situations that generally preclude ET‐CO2 monitoring such as high frequency ventilation, apnea testing, and noninvasive ventilation. TC‐CO2 monitoring has also been used in spontaneously breathing children with airway and respiratory issues such as croup and status asthmaticus as well as to monitor metabolic status during treatment of acidosis related to diabetic ketoacidosis.Conclusions: Transcutaneous carbon dioxide monitoring may be a useful adjunct in various clinical scenarios in infants and children. It should be viewed as a complimentary technology and may be used in combination with ET‐CO2 monitoring.Keywords
This publication has 47 references indexed in Scilit:
- Noninvasive monitoring of PaCO2 during one-lung ventilation and minimal access surgery in adults: End-tidal versus transcutaneous techniquesJournal of Minimal Access Surgery, 2007
- A Comparative Evaluation of Transcutaneous and End-Tidal Measurements of CO2 in Thoracic AnesthesiaAnesthesia & Analgesia, 2003
- A Combined Ear Sensor for Pulse Oximetry and Carbon Dioxide Tension Monitoring: Accuracy in Critically Ill ChildrenAnesthesia & Analgesia, 2003
- Transcutaneous Carbon Dioxide Measurement During Apnea Testing in Pediatric PatientsJournal of Intensive Care Medicine, 2001
- Noninvasive monitoring of end-tidal CO2 via nasal cannulas in spontaneously breathing children during the perioperative periodCritical Care Medicine, 1994
- Transcutaneous and arterial blood gas monitoring during acute asthmatic symptoms in older childrenPediatric Pulmonology, 1992
- Transcutaneous carbon dioxide pressure for monitoring patients with acute upper airway obstructionThe Journal of Pediatrics, 1991
- Noninvasive monitoring of carbon dioxide: A comparison of the partial pressure of transcutaneous and end-tidal carbon dioxide with the partial pressure of arterial carbon dioxideJournal of Clinical Monitoring and Computing, 1987
- End-tidal Pco2 Monitoring in Infants and Children Ventilated with Either a Partial Rebreathing or a Non-rebreathing CircuitAnesthesiology, 1987
- Transcutaneous monitoring in the newborn infantThe Journal of Pediatrics, 1983