Optimizing Portable Pulse Oximeter Measurement Accuracy and Consistency During Exercise
Authors:
- Jonas LaPierJLJonas LaPier
- Megan ChatellierMCMegan Chatellier
- 1 July 2017
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Acute Care Physical Therapy
- Vol. 8 (3) , 96-105
Abstract
Monitoring oxygen saturation (SpO2) in some patient populations during exercise is vital. The objectives of this study were to determine whether under conditions of artificial hypoperfusion pulse oximeter performance during exercise was influenced by sensor type, attachment clip reinforcement, or a low pressure arm cuff. Six different portable oximeter models were tested on study participants at rest with a control (normothermic) and experimental (hypothermic) hand. Oximeters were then tested during treadmill walking at 1 and 2 mph with the experimental hand remaining hypothermic. Electrocardiographic measurements of heart rate were also taken simultaneously. There were differences in SpO2 values between oximeter pairs following cold exposure at rest for most models (5 out of 6) but during walking for only 1 model. There were also differences between oximeter and electrocardiographic heart rate measurements for some models when study participants walked on a treadmill at 1 mph (1 out of 6) and 2 mph (3 out of 6). Results suggest that oximeter performance is better when measuring SpO2 than heart rate. Cold-induced hypoperfusion caused inaccuracies in SpO2 measurements, which were greater at rest than during exercise. Study results also suggest that some pulse oximeters perform better when the oximeter sensor attachment clip is reinforced with an elastic band.Keywords
This publication has 29 references indexed in Scilit:
- Photoplethysmographic signals and blood oxygen saturation values during artificial hypothermia in healthy volunteersPhysiological Measurement, 2012
- Performance of three new-generation pulse oximeters during motion and low perfusion in volunteersJournal of Clinical Anesthesia, 2012
- Air travel and chronic obstructive pulmonary disease: a new algorithm for pre-flight evaluationThorax, 2012
- Comparison of oxygen saturation values and measurement times by pulse oximetry in various parts of the bodyApplied Nursing Research, 2011
- The accuracy of pulse oximetry in emergency department patients with severe sepsis and septic shock: a retrospective cohort studyBMC Emergency Medicine, 2010
- Automatic noninvasive measurement of systolic blood pressure using photoplethysmographyBioMedical Engineering OnLine, 2009
- Dark Skin Decreases the Accuracy of Pulse Oximeters at Low Oxygen Saturation: The Effects of Oximeter Probe Type and GenderAnesthesia & Analgesia, 2007
- The Effects of Motion on the Performance of Pulse Oximeters in Volunteers (Revised publication)Anesthesiology, 1997
- Comparison of Oxygen Saturation by Pulse Oximetry and Co-oximetry During Exercise Testing in Patients With COPDChest, 1996
- Electrical surface conductivity in quartz induced by ion implantationJournal of Applied Physics, 1992