Abstract
Chrysin is a flavonoid with many biological and medicinal activities, such as anti-cancer, anti-inflammatory, anti-diabetic, and antioxidant properties, and has medicinal value. Based on the Quality by Design approach, the present study used a UV-spectrophotometer to develop a bioanalytical method for chrysin in human plasma. The validation of the method was performed as per ICH M10 guidelines. The technique was performed using methanol as a solvent and the spectrum was recorded at a wavelength of 370 nm. The spiked chrysin from plasma was extracted using the protein precipitation method. The developed method was linear with an r2 value of 0.9996 over a concentration range of 4-20 μg/ml. The results of all validation parameters are found to be within the accepted limits with a % RSD value of less than 2% and the % recovery was greater than 95%. The % RSD and % recovery results confirmed that the method was precise and accurate in the study. LOD and LOQ values in plasma samples were found to be 0.47 μg/ml and 1.43 μg/ml respectively. The stability studies like freeze-thaw, and short-term stability studies were also performed and proved the reliability of the method. Therefore, the developed bioanalytical method can effectively estimate chrysin in plasma samples within its sensitivity limits.