Contrasting Effects of Convective Flow on Catheter Ablation Lesion Size: Cryo Versus Radiofrequency Energy
- 28 February 2008
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 31 (3) , 300-307
Abstract
Background: Cryoablation has now become an alternative to treat many cardiac arrhythmias, and may be the treatment of choice in some patient populations. We compared the effects of convective flow on large‐tip cryo and radiofrequency (RF) lesions dimensions. Methods: Cryoablation and RF ablation were performed on porcine heart sections in a saline bath with varying directed flow rates. Cryoablation was performed for 4 minutes on 50 tissue pieces with tip temperature controlled at −80°C. RF ablation was performed on 50 tissue pieces for 60 seconds at 60°C tip temperature. The pieces were placed in culture media for 24 hours, and then sectioned, stained, and measured. Results: Cryoablation and RF lesion sizes varied significantly with flow such that higher flow rates produced smaller cryoablation lesions and larger RF lesions (mean cryoablation volumes: 854 ± 402, 808 ± 217, 781 ± 217, 359 ± 114, and 292 ± 117 mm3, and mean RF volumes: 211 ± 35, 304 ± 79, 439 ± 125, 525 ± 187, and 597 ± 126 mm3 for 0, 1, 2, 3, and 5L/min flow rates, respectively, P < 0.0005). Trabeculated pieces had larger cryoablation lesions and smaller RF lesions than nontrabeculated ones at higher flow rate (P < 0.005). Cryoablation lesion volume increased as the time to reach −80°C decreased (r2= 0.72). Conclusion: In contrast to RF ablation, cryoablation lesion size is smaller at high flow rates, and larger at low flow rates due to the warming effects of local convective flow. The effects of high flow are reduced in areas of trabeculation, and the time to reach −80°C predicts cryoablation lesion size.Keywords
This publication has 24 references indexed in Scilit:
- Determinants of Lesion Sizes and Tissue Temperatures During Catheter CryoablationPacing and Clinical Electrophysiology, 2007
- Convective Cooling Effect on Cooled‐Tip Catheter Compared to Large‐Tip Catheter Radiofrequency AblationPacing and Clinical Electrophysiology, 2006
- Cryoablation versus radiofrequency ablation for treatment of pediatric atrioventricular nodal reentrant tachycardia: Initial experience with 4-mm cryocatheterHeart Rhythm, 2006
- Catheter ablation for atrioventricular nodal reentry tachycardia in children: A time to freeze, and a time to burnHeart Rhythm, 2006
- Cryoablation for accessory pathways located near normal conduction tissues or within the coronary venous system in children and young adultsHeart Rhythm, 2006
- Transcatheter cryoablation of tachyarrhythmias in childrenJACC, 2005
- Cryoenergy Catheter Ablation:Journal of Cardiovascular Electrophysiology, 2004
- In-vivo measurement of swine myocardial resistivityIEEE Transactions on Biomedical Engineering, 2002
- Guidelines for predicting lesion size at common endocardial locations during radio-frequency ablationIEEE Transactions on Biomedical Engineering, 2001
- Temperature‐Controlled Irrigated Tip Radiofrequency Catheter Ablation:Journal of Cardiovascular Electrophysiology, 1998