Design and analysis of a continuum robot for transnasal skull base surgery
- 31 August 2021
- journal article
- research article
- Published by Wiley in The International Journal of Medical Robotics and Computer Assisted Surgery
- Vol. 17 (6) , e2328
Abstract
Background: The traditional surgical instruments for transnasal endoscopic skull base surgery have poor flexibility and continuum robot has attracted extensive attention for its high dexterity and safety.Methods: A new type of micro continuum surgical robot is designed. On the basis of kinematic analysis, a superposition method is proposed to analyse the reachable configuration space of the system, and the workspace is analysed. The results obtained by superposition method can provide reference for the selection of joint parameters. At last, the static model of the system is established considering friction and coupling between two segments.Results: The simulation results show that the workspace can meet the requirements of surgery. And the validity of the static model is verified by numerical simulation and experiment, which lays a foundation for the establishment of the driving force transmission system and precise control of the robot.Conclusions: The research results of this study contribute to the real‐time control and movement of robot. The proposed continuum robot provides convenient conditions for the clinical application.Keywords
Funding Information
- Natural Science Foundation of Jiangsu Province
- Natural Science Foundation of Jiangsu Province (BK20191272)
- Fundamental Research Funds for the Central Universities (NS2020036)
This publication has 29 references indexed in Scilit:
- Development of a compact continuum tubular robotic system for nasopharyngeal biopsyMedical & Biological Engineering & Computing, 2016
- Development of a Multi-Channel Concentric Tube Robotic System With Active Vision for Transnasal Nasopharyngeal Carcinoma ProceduresIEEE Robotics and Automation Letters, 2016
- A Novel Continuum Robot Using Twin-Pivot Compliant Joints: Design, Modeling, and ValidationJournal of Mechanisms and Robotics, 2015
- Tendon-driven continuum robot for neuroendoscopy: validation of extended kinematic mapping for hysteresis operationInternational Journal of Computer Assisted Radiology and Surgery, 2015
- An Experimental Feasibility Study on Robotic Endonasal TelesurgeryNeurosurgery, 2015
- Design and analysis of a family of snake arm robots connected by compliant jointsMechanism and Machine Theory, 2014
- Compliant articulation structure using superelastic NiTiNOLSmart Materials and Structures, 2013
- Innovative endoscopic sino-nasal and anterior skull base roboticsInternational Journal of Computer Assisted Radiology and Surgery, 2013
- Clinical Requirements and Possible Applications of Robot Assisted Endoscopy in Skull Base and Sinus Surgery2010
- Design and Integration of a Telerobotic System for Minimally Invasive Surgery of the ThroatThe International Journal of Robotics Research, 2009