Miniaturized Fontan Circulation Assist Device: Chronic In Vivo Evaluation
- 19 April 2021
- journal article
- research article
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 67 (11) , 1240-1249
Abstract
We have miniaturized and optimized our implantable rotary blood pump developed to provide long-term mechanical right heart support for patients who have failing Fontan circulation. The objective of this study was to evaluate the miniaturized Fontan circulation assist device (mini-FCAD) during 30-day sheep studies (n = 5). A complete right heart bypass was performed and all return flow was supported by the pump. Postoperatively, unfractionated heparin was given to maintain thromboelastography R times of 2× normal. The first two studies were terminated on day 0 and day 4 due to complications. In the final three studies, the animals remained healthy and were electively terminated at 30 ± 2 days. Pump flow was between 5 and 7 lpm, left atrial pressure remained normal, and inlet pressures were between 3 and 18 mm Hg with no incidents of suction. There was no evidence of hemolysis, end organ or pulmonary dysfunction, thromboembolic events, nor thermal damage to the surrounding tissue. Explanted devices from two studies were free of thrombi and in the third study there were unattached thrombi on the SVC inlet of the rotor. The mini-FCAD was successfully tested in vivo as a right heart replacement device demonstrating adequate circulatory support and normal physiologic pulmonary and venous pressures.Keywords
This publication has 18 references indexed in Scilit:
- Validating Left Ventricular Filling Pressure Measurements in Patients with Congestive Heart Failure: CardioMEMS™ Pulmonary Arterial Diastolic Pressure versus Left Atrial Pressure Measurement by Transthoracic EchocardiographyCardiology Research and Practice, 2018
- State of the art of the Fontan strategy for treatment of univentricular heart diseaseF1000Research, 2018
- Two-Year Outcomes with a Magnetically Levitated Cardiac Pump in Heart FailureNew England Journal of Medicine, 2018
- Fontan Failure and Death in Contemporary Fontan Circulation: Analysis From the Last Two DecadesThe Annals of Thoracic Surgery, 2018
- Cavopulmonary mechanical circulatory support in Fontan patients and the need for physiologic control: A computational study with a closed-loop exercise modelThe International Journal of Artificial Organs, 2018
- Ventricular assist device use in single ventricle congenital heart diseasePediatric Transplantation, 2017
- Left Atrial Pressure Monitoring With an Implantable Wireless Pressure Sensor After Implantation of a Left Ventricular Assist DeviceASAIO Journal, 2017
- New aspects on efficient anticoagulation and antiplatelet strategies in sheepBMC Veterinary Research, 2013
- Blood flow distribution in a large series of patients having the Fontan operation: A cardiac magnetic resonance velocity mapping studyThe Journal of Thoracic and Cardiovascular Surgery, 2009
- Anticoagulation of Juvenile Sheep and Goats With Heparin, Warfarin, and ClopidogrelASAIO Journal, 2007