Determination of the Sample Thicknesses by Intensity Ratio Measurement by Energy Dispersive X-Ray Fluorescence Spectrometry
Authors:
- 1 June 1999
- journal article
- research article
- Published by Taylor & Francis in Instrumentation Science & Technology
- Vol. 27 (3) , 181-189
Abstract
A procedure for the determination of the thickness of Au and Ag films and foils coated on a metallic Cu foil (substrate) is described. The procedure is based on the measurements of the fluorescence intensity ratios for two lines from the substrate element, excited by 59.5 keV photons emitted from a filtered Am-241 radioisotope source. The result of the procedure yielded between 0.039×10−5 and 4.3322×10−5 g/cm2 for Au and between 0.0394×10−5 and 5.8364×10−5 g/cm2 for Ag. These are tested by gravimetric measurements. The proposed procedure can also be employed to determine the density of the covering material or the incoming and outgoing angles of the radiation if the film thickness is known.Keywords