Development of a Polydimethylsiloxane‐Based Flexible Digital Bolus Dosimeter: From Unit‐Cell Characterization to Linear Array for Skin‐Dose Distribution Verification
Open Access
- 30 July 2026
- journal article
- research article
- Published by Wiley in Physica Status Solidi (a)
- Vol. 223 (15) , e70451
Abstract
In this study, we developed a flexible digital bolus dosimeter based on a polycrystalline HgI 2 photoconductor embedded in a tissue‐equivalent polydimethylsiloxane (PDMS) matrix, incorporating a 70 μm protective layer for skin‐dose evaluation. The dosimeter exhibited excellent linearity ( R 2 = 0.9996) within 2–400 MU and superior repeatability (standard deviation of 1.59%). Although the signal gradually decreased by approximately 8.3% (range from 5.6% to −2.7%) as the dose rate increased from 100 to 600 MU/min due to charge trapping and polarization, a stable response was maintained using a dose‐rate‐dependent correction factor or integrating advanced signal processing techniques. Mechanical stability tests up to 1000 bending cycles suggested the need for baseline correction on curved surfaces due to increased drift‐charge and photo‐charge. Subsequently, a 24‐pixel linear array dosimeter was fabricated on a flexible printed circuit board (fPCB) to evaluate continuous spatial dose distribution. When applied to a curved RSD chest phantom, the array reproduced consistent spatial dose distribution trends compared to radiochromic film at various gantry angles, with absolute dose differences of 3.51–7.05 cGy (7%–12%). These results demonstrate that the proposed platform is practical and suitable for evaluating skin‐dose distributions in complex geometries during radiotherapy.Keywords
This publication has 24 references indexed in Scilit:
- The synchronous detection technique for the accurate monitoring of high-energy pulsed X-raysNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2023
- Dosimetric considerations for moldable silicone composites used in radiotherapy applicationsJournal of Applied Clinical Medical Physics, 2022
- Technical note: Evaluation of a silicone‐based custom bolus for radiation therapy of a superficial pelvic tumorJournal of Applied Clinical Medical Physics, 2022
- Modeling Radiotherapy Side Effects: Practical Applications for Planning Optimization. RancatiTiziana, FiorinoClaudio. Series in Medical Physics and Biomedical Engineering. Series Editors: Webster John G, Russell Ritenour E, Tabakov Slavik, Ng Kwan‐Hoong Boca Raton, FL: CRC Press, Routledge Taylor & Francis Group, 2019. Hardcover: 494 pp. Price: $168.00. ISBN: 9781315270814Medical Physics, 2021
- Radiological tissue equivalence of deformable silicone‐based chemical radiation dosimeters (FlexyDos3D)Journal of Applied Clinical Medical Physics, 2019
- An optimized calibration method for surface measurements with MOSFETs in shaped-beam radiosurgeryPhysica Medica, 2014
- Radiochromic film dosimetry: Past, present, and futurePhysica Medica, 2011
- In Vivo Dosimetry with MOSFET Detector during RadiotherapyIFMBE Proceedings, 2007
- Radiation dermatitis: Clinical presentation, pathophysiology, and treatment 2006Journal of the American Academy of Dermatology, 2006
- Comparison of PbI2 and HgI2 for direct detection active matrix x-ray image sensorsJournal of Applied Physics, 2002